Wireless integrated circuit for 100-channel charge-balanced neural stimulation

Update Item Information
Publication Type Journal Article
School or College College of Engineering
Department Electrical & Computer Engineering
Creator Harrison, Reid R.
Other Author Thurgood, Brandon K.; Warren, David J.; Ledbetter, Noah M.; Clark, Gregory A.
Title Wireless integrated circuit for 100-channel charge-balanced neural stimulation
Date 2009-12
Description The authors present the design of an integrated circuit for wireless neural stimulation, along with benchtop and in-vivo experimental results. The chip has the ability to drive 100 individual stimulation electrodes with constant-current pulses of varying amplitude, duration, interphasic delay, and repetition rate. The stimulation is performed by using a biphasic (cathodic and anodic) current source, injecting and retracting charge from the nervous system. Wireless communication and power are delivered over a 2.765-MHz inductive link. Only three off-chip components are needed to operate the stimulator: a 10-nF capacitor to aid in power-supply regulation, a small capacitor (100 pF) for tuning the coil to resonance, and a coil for power and command reception. The chip was fabricated in a commercially available 0.6- m 2P3M BiCMOS process. The chip was able to activate motor fibers to produce muscle twitches via a Utah Slanted Electrode Array implanted in cat sciatic nerve, and to activate sensory fibers to recruit evoked potentials in somatosensory cortex.
Type Text
Publisher Institute of Electrical and Electronics Engineers (IEEE)
Journal Title IEEE Transactions on Biomedical Circuits and Systems
Volume 3
Issue 6
First Page 405
Last Page 414
DOI 10.1109/TBCAS.2009.2032268
citatation_issn 1932-4545
Subject Wireless integrated circuit; Low-power
Subject LCSH Integrated circuits; Neural stimulation; Microelectrodes; Implants, Artificial
Language eng
Bibliographic Citation Thurgood, B. K., Warren, D. J., Ledbetter, N. M., Clark, G. A., & Harrison, R. R. (2009). Wireless integrated circuit for 100- channel charge-balanced neural stimulation. IEEE Transactions on Biomedical Circuits and Systems, 3(6), 405-14.
Rights Management (c) 2009 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.