Page 17

Update Item Information
Title A Differential Ultraviolet Resonant Absorption System for the Non-Intrusive Measurement of Nitric Oxide in a Combustion Environment
Creator Murphree, D. L.; Srikantaiah, D. V.; Stickel, R. E.; Daubach, R. O.
Publisher Digitized by J. Willard Marriott Library, University of Utah
Date 1983
Spatial Coverage Akron, Ohio
Abstract An Instrument has been constructed based upon the differential ultraviolet resonant absorption detection of nitric oxide. A hollow cathode lamp produces strong gamma band emission of nitric oxide which is spectrally modified to generate a system response function which is solely sensitive to nitric oxide concentration. The system has been calibrated from 305 K to 1800 K. The rapid response of the system and its immunity to interferences associated with intrusive probe measurements make it a particularly favorable means for determining nitric oxide concentration in harsh environments.
Type Text
Format application/pdf
Language eng
Rights This material may be protected by copyright. Permission required for use in any form. For further information please contact the American Flame Research Committee.
Conversion Specifications Original scanned with Canon EOS-1Ds Mark II, 16.7 megapixel digital camera and saved as 400 ppi uncompressed TIFF, 16 bit depth.
Scanning Technician Cliodhna Davis
ARK ark:/87278/s6v127c6
Setname uu_afrc
ID 635
Reference URL https://collections.lib.utah.edu/ark:/87278/s6v127c6