Publication Type |
Journal Article |
School or College |
College of Engineering |
Department |
Electrical & Computer Engineering |
Creator |
Harrison, Reid R. |
Other Author |
Thurgood, B. K.; Ledbetter, N. M.; Warren, D. J.; Clark, G. A. |
Title |
Wireless integrated circuit for 100-channel neural stimulation |
Date |
2008-11 |
Description |
We present the design of an integrated circuit for wireless neural stimulation, along with bench-top and in-vivo experimental results. The chip has the ability to drive 100 individual stimulation electrodes with constant-current pulses of varying amplitude, duration, interphasic delay, and repetition rate. The stimulation is done using a biphasic (cathodic and anodic) current source, injecting and retracting charge from the nervous system. Wireless communication and power are achieved over a 2.765-MHz inductive link. Only two off-chip components are needed to operate the stimulator: a 10-nF capacitor to aid in power supply regulation and a coil for power and command reception. The chip was fabricated in a commercially available 0.6-μm 2P3M BiCMOS process. The chip was able to activate motor fibers to produce muscle twitches via a Utah Slanted Electrode Array implanted in cat sciatic nerve, and to activate sensory fibers to recruit evoked potentials in somatosensory cortex. |
Type |
Text |
Publisher |
Institute of Electrical and Electronics Engineers (IEEE) |
First Page |
129 |
Last Page |
132 |
DOI |
10.1109/BIOCAS.2008.4696891 |
Subject |
Wireless; Integrated Neural Interface; Neural recording; Telemetry |
Subject LCSH |
Integrated circuits; Microelectrodes; Implants, Artificial; Biotelemetry; Neural stimulation; Cats as laboratory animals; Animal models in research |
Language |
eng |
Conference Title |
2008 IEEE Biomedical Circuits and Systems Conference; ; Baltimore, MD, USA |
Bibliographic Citation |
Thurgood, B. K., Ledbetter, N. M., Warren, D. J., Clark, G. A., & Harrison, R. R. (2008). Wireless integrated circuit for 100- channel neural stimulation. Proceedings of the 2008 IEEE Biomedical Circuits and Systems Conference (BioCAS 2008), 129-32. |
Rights Management |
(c) 2008 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE. |
Format Medium |
application/pdf |
Format Extent |
461,657 bytes |
Identifier |
ir-main,14024 |
ARK |
ark:/87278/s68d0dtv |
Setname |
ir_uspace |
ID |
706478 |
Reference URL |
https://collections.lib.utah.edu/ark:/87278/s68d0dtv |