Publication Type |
Journal Article |
School or College |
College of Science |
Department |
Physics |
Creator |
Symko, Orest George |
Other Author |
Ribeiro, P. Costa; Bruno, A. C.; Paulsen, C. C. |
Title |
Spatial Fourier transform method for evaluating SQUID gradiometers |
Date |
1987 |
Description |
A simple method of measuring the spatial transfer function of a gradiometer, consisting of a flux transformer coupled to a SQUID, is presented and it is compared with theoretical predictions. Based, on this approach, a new method of reporting a gradiometer's performance is proposed; the rejection factor is expressed in decibels obtained directly from the transfer function plot. |
Type |
Text |
Publisher |
American Institute of Physics (AIP) |
Volume |
58 |
Issue |
8 |
First Page |
1510 |
Last Page |
1513 |
DOI |
10.1063/1.1139390 |
Subject |
Spatial Fourier transform; SQUID gradiometers; Evaluation; Spatial transfer function |
Subject LCSH |
Superconducting quantum interference devices; Biomagnetism -- Measurement |
Language |
eng |
Bibliographic Citation |
Ribeiro, P. C., Bruno, A. C., Paulsen, C. C., & Symko, O. G. (1987). Spatial Fourier transform method for evaluating SQUID gradiometers. Review of Scientific Instruments, 58(8), 1510-3. |
Rights Management |
©American Institute of Physics. The following article appeared in Ribeiro, P. C., Bruno, A. C., Paulsen, C. C., & Symko, O. G., Review of Scientific Instruments, 58(8), 1987 and may be found at http://dx.doi.org/10.1063/1.1139390 |
Format Medium |