Publication Type |
Journal Article |
School or College |
School of Medicine |
Department |
Neurosurgery |
Creator |
Kestle, John R. W. |
Other Author |
Drake, James M.; Milner, Ruth; Cinalli, Giuseppe; Boop, Frederick; Piatt, Joseph Jr.; Haines, Stephen; Schiff, Steven J.; Cochrane, D. Douglas; Steinbok, Paul; MacNeil, Nancy |
Title |
Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus |
Date |
1998 |
Description |
OBJECTIVE: Forty percent of standard cerebrospinal fluid shunts implanted for the treatment of pediatric hydrocephalus fail within the first year. Two new shunt valves designed to limit excess flow, particularly in upright positions, were studied to compare treatment failure rates with those for standard differential-pressure valves. METHODS: Three hundred-forty-four hydrocephalic children (age, birth to 18 yr) undergoing their first cerebrospinal fluid shunt insertion were randomized at 12 North American or European pediatric neurosurgical centers. Patients received one of three valves, i.e., a standard differential-pressure valve; a Delta valve (Medtronic PS Medical, Goleta, CA), which contains a siphon-control component designed to reduce siphoning in upright positions; or an Orbis-Sigma valve (Cordis, Miami, FL), with a variable-resistance, flow-limiting component. Patients were monitored for a minimum of 1 year. Endpoints were defined as shunt failure resulting from shunt obstruction, overdrainage, loculations of the cerebral ventricles, or infection. Outcome events were assessed by blinded independent case review. RESULTS: One hundred-fifty patients reached an endpoint; shunt obstruction occurred in 108 (31.4%), overdrainage in 12(3.5%), loculated ventricles in 2(0.6%), and infection in 28 (8.1%). Sixty-one percent were shunt failure-free at 1 year and 47% at 2 years, with a median shunt failure-free duration of 656 days. There was no difference in shunt failure-free duration among the three valves (P = 0.24). |
Type |
Text |
Publisher |
Wolters Kluwer (LWW) |
Volume |
43 |
Issue |
2 |
First Page |
294 |
Last Page |
303 |
Subject |
Shunt failure; Valve design; Orbis-Sigma valve; Delta valve |
Subject LCSH |
Cerebrospinal fluid shunts; Hydrocephalus in children |
Language |
eng |
Bibliographic Citation |
Drake, J. M., Kestle, J. R. W, Milner, R., Cinalli, G., Boop, F., Piatt, J. Jr., Haines, S., Schiff, S. J., Cochrane, D. D., Steinbok, P., & MacNeil, N. (1998). Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus. Neurosurgery, 43(2), 294-303. |
Rights Management |
(c) Wolters Kluwer (LWW) http://lww.com |
Format Medium |
application/pdf |
Format Extent |
1,048,744 bytes |
Identifier |
ir-main,13169 |
ARK |
ark:/87278/s6mg870v |
Setname |
ir_uspace |
ID |
705747 |
Reference URL |
https://collections.lib.utah.edu/ark:/87278/s6mg870v |