Description This resource features a photograph of a Line Isolation Monitor. Isolated power systems are frequently used in Operating Rooms and Intensive Care Units. Such systems use an isolation transformer system so that neither of the two output lines powering the equipment offers any voltage with respect to ground. This helps eliminate the shock hazard associated with working in potentially wet clinical environments.