| Title | Contact force analysis under femoral head micro-separation for ceramic-on-ceramic hip implants: finite element analysis and experimental validation |
| Publication Type | thesis |
| School or College | College of Engineering |
| Department | Mechanical Engineering |
| Author | Kakarla, Lakshmi Navya Deepika |
| Date | 2012-05 |
| Description | When a prosthetic hip fails to perform like an ideal ball-and-socket joint but instead permits small subluxation and edge loading between the joint members, the bearing surfaces are subjected to increased contact stresses and wear that could ultimately cause failure and require costly and painful revision of the prosthesis. The objectives of the present study were (1) to model and analyze one such adverse motion: rapid femoral head reduction; (2) to quantify the ensuing dynamic contact force and elevated contact stresses; and (3) to improve the relevance of hip joint wear tests by providing validated contact force values representing worst case conditions in ceramic-on-ceramic hip bearings. A dynamic model of head micro-separation occurring during normal human gait was examined via a combined approach of laboratory testing and finite element analysis (FEA). The testing validated the FEA (though with some explainable error) against measured values for normal velocity at a point on the femoral head and strain on the femur and femoral neck. Then, the FE model was used to analyze the contact forces and the stresses during the edge loading. This approach contradicted a key hypothesis of the study, specifically, that the duration of edge loading contact would be close to the period of vibrations in the femur. It was revealed that the peak contact stresses are strongly influenced by the model's input conditions rather than the femur's natural vibration characteristics. The results showed an increase in the peak femoral head velocity and the peak femoral strains during the reduction event as the micro-separation increased. Validation of the FEA against experimental results showed large errors in velocity and relatively small errors in strain. The submodel analysis showed much higher contact force and contact stress than the global model analysis, which was attributed to boundary conditions (BCs) and limitations of the software used. A better understanding of the cause of the errors in the approach undertaken demands an improvement in the submodeling procedure, which involves a more detailed analysis beyond the scope of this study. |
| Type | Text |
| Publisher | University of Utah |
| Subject | contact force; femoral head; micro-separation; microseparation; ceramic; hip implants; contact stresses |
| Subject LCSH | Artificial hip joints -- Testing; Artificial hip joints -- Materials |
| Dissertation Institution | University of Utah |
| Dissertation Name | Master of Science |
| Language | eng |
| Rights Management | © Lakshmi Navya Deepika Kakarla |
| Format | application/pdf |
| Format Medium | application/pdf |
| Format Extent | 4,798,006 bytes |
| Identifier | us-etd3/id/673 |
| Source | Original in Marriott Library Special Collections, RD14.5 2012 .K37 |
| ARK | ark:/87278/s6qf97nb |
| DOI | https://doi.org/doi:10.26053/0H-32N8-PJG0 |
| Setname | ir_etd |
| ID | 194837 |
| OCR Text | Show CONTACT FORCE ANALYSIS UNDER FEMORAL HEAD MICRO-SEPARATION FOR CERAMIC-ON-CERAMIC HIP IMPLANTS: FINITE ELEMENT ANALYSIS AND EXPERIMENTAL VALIDATION by Lakshmi Navya Deepika Kakarla A thesis submitted to the faculty of The University of Utah in partial fulfillment of the requirements for the degree of Master of Science Department of Mechanical Engineering The University of Utah May 2012 Copyright © Lakshmi Navya Deepika Kakarla 2012 All Rights Reserved Th e Uni v e r s i t y o f Ut ah Gr ad ua t e S ch o o l STATEMENT OF THESIS APPROVAL The thesis of Lakshmi Navya Deepika Kakarla has been approved by the following supervisory committee members: Rebecca M. Brannon , Chair 2/27/2012 Date Approved Daniel O. Adams , Member 2/27/2012 Date Approved Kenneth L. Monson , Member 2/27/2012 Date Approved and by Timothy Ameel , Chair of the Department of Mechanical Engineering and by Charles A. Wight, Dean of The Graduate School. ABSTRACT When a prosthetic hip fails to perform like an ideal ball-and-socket joint but instead permits small subluxation and edge loading between the joint members, the bearing surfaces are subjected to increased contact stresses and wear that could ultimately cause failure and require costly and painful revision of the prosthesis. The objectives of the present study were (1) to model and analyze one such adverse motion: rapid femoral head reduction; (2) to quantify the ensuing dynamic contact force and elevated contact stresses; and (3) to improve the relevance of hip joint wear tests by providing validated contact force values representing worst case conditions in ceramic-on-ceramic hip bearings. A dynamic model of head micro-separation occurring during normal human gait was examined via a combined approach of laboratory testing and finite element analysis (FEA). The testing validated the FEA (though with some explainable error) against measured values for normal velocity at a point on the femoral head and strain on the femur and femoral neck. Then, the FE model was used to analyze the contact forces and the stresses during the edge loading. This approach contradicted a key hypothesis of the study, specifically, that the duration of edge loading contact would be close to the period of vibrations in the femur. It was revealed that the peak contact stresses are strongly influenced by the model's input conditions rather than the femur's natural vibration characteristics. iv The results showed an increase in the peak femoral head velocity and the peak femoral strains during the reduction event as the micro-separation increased. Validation of the FEA against experimental results showed large errors in velocity and relatively small errors in strain. The submodel analysis showed much higher contact force and contact stress than the global model analysis, which was attributed to boundary conditions (BCs) and limitations of the software used. A better understanding of the cause of the errors in the approach undertaken demands an improvement in the submodeling procedure, which involves a more detailed analysis beyond the scope of this study. TABLE OF CONTENTS ABSTRACT....................................................................................................................... iii LIST OF TABLES........................................................................................................... viii LIST OF ABBREVIATIONS............................................................................................ ix ACKNOWLEDGEMENTS................................................................................................ x CHAPTER 1. INTRODUCTION ........................................................................................................ 11 1.1 Total Hip Arthroplasty........................................................................................ 11 1.2 Implant Materials ................................................................................................ 11 1.3 Background ........................................................................................................... 2 1.3.1 COC Implants - Squeaking - Stripe Wear .................................................... 2 1.3.2 Micro-separation / Micro-lateralization......................................................... 4 1.3.3 Contact Force between Implants.................................................................... 6 1.4 Motivation............................................................................................................. 8 1.5 Objectives ........................................................................................................... 10 1.6 Specific and Novel Contributions....................................................................... 12 2. BIOMECHANICAL REDUCTION (BMR) MODEL................................................. 15 2.1 Introduction......................................................................................................... 15 2.2 Components of the BMR Model......................................................................... 16 2.3 BMR Model Configuration................................................................................. 18 2.3.1 Standard Orientations................................................................................... 19 2.3.2 Feasible Setup Orientations ......................................................................... 19 2.3.3 Final Configuration...................................................................................... 20 2.4 Initial Conditions ................................................................................................ 20 2.5 Inputs and Boundary Conditions ........................................................................ 20 2.6 Rigid Body Analysis of BMR Model ................................................................. 22 2.6.1 Equations of Motion .................................................................................... 23 2.6.2 Effect of the Spring Force on the Contact Force ......................................... 24 vi 3. ELEMENTARY ANALYSIS AND TESTING ........................................................... 33 3.1 Introduction......................................................................................................... 33 3.2 Problem Statement .............................................................................................. 34 3.3 Methods............................................................................................................... 34 3.3.1 Laboratory testing ........................................................................................ 34 3.3.2 Axisymmetric Finite Element Modeling ..................................................... 36 3.3.3 3D Finite Element Analysis ......................................................................... 36 3.4 Results: Comparison of Analytical, Experimental and Axisymmetric FEA ...... 37 3.5 Conclusion .......................................................................................................... 38 4. LABORATORY TESTING OF BMR MODEL .......................................................... 44 4.1 Introduction......................................................................................................... 44 4.2 The Femur Construct .......................................................................................... 44 4.3 Fixtures and Initial Alignment ............................................................................ 46 4.4 Spring Mechanism.............................................................................................. 48 4.5 Laser Vibrometer ................................................................................................ 48 4.6 Strain Gaging ...................................................................................................... 49 4.7 Data Acquisition ................................................................................................. 50 4.8 Experimental Procedures .................................................................................... 50 4.8.1 Input Velocity .............................................................................................. 50 4.8.2 Reduction Trials........................................................................................... 51 5. FINITE ELEMENT ANALYSIS OF BMR MODEL .................................................. 59 5.1 Introduction......................................................................................................... 59 5.2 Three Dimensional Solid Model and Mesh Generation ..................................... 60 5.3 Material Properties.............................................................................................. 62 5.4 Loading and Boundary Conditions ..................................................................... 64 5.5 Contact Modeling................................................................................................ 65 5.6 Submodeling of BMR Model.............................................................................. 66 5.6.1 Submodel Generation................................................................................... 67 5.6.2 Boundary Conditions ................................................................................... 67 5.6.3 Submodeling Analysis Procedure ................................................................ 67 6. RESULTS: TESTING AND ANALYSIS OF BMR MODEL..................................... 75 6.1 Experimental Results .......................................................................................... 75 6.1.1 Velocity of the Femoral Head...................................................................... 75 6.1.2 Strain in the Femur and the Femoral Neck .................................................. 76 6.2 Validation of the BMR Model ............................................................................ 76 6.2.1 Velocity of the Femoral Head...................................................................... 76 6.2.2 Strain in the Femur and the Femoral Neck .................................................. 78 6.3 Inferred Outputs from Global Model and Submodel Analyses .......................... 78 6.4 Contour Plots ...................................................................................................... 79 vii 7. DISCUSSION............................................................................................................... 91 7.1 Validation of the PSG in the FE Model .............................................................. 91 7.2 Behavior of the Femoral Cortex Material ........................................................... 92 7.3 Effect of Friction................................................................................................. 93 7.3.1 Effect of Lubrication on the PSG................................................................. 93 7.3.2 Effect of Friction on the Velocity of the Femoral Head .............................. 94 7.3.3 Effect of Friction on the Contact Force ....................................................... 95 7.4 Key Hypothesis ................................................................................................... 96 7.5 Submodeling Procedure - Future Scope............................................................. 97 7.5.1 Contact Stiffness - Boundary Conditions.................................................... 98 7.5.2 Successive Submodeling Approach to Improve Consistency and Convergence ......................................................................................................... 99 8. CONCLUSIONS AND FUTURE SCOPE................................................................. 103 8.1 Conclusions....................................................................................................... 103 8.2 Future Scope ..................................................................................................... 104 APPENDIX A. BMR MODEL CONFIGURATION PROCEDURE 110 B. FEMUR-IMPLANT ASSEMBLY AND FEMUR CONSTRUCT MEASUREMENTS 113 REFERENCES ............................................................................................................... 116 LIST OF TABLES 1: Available prosthetic hip material couples and their corresponding acronyms ............. 13 2: Description of the symbols used in the FBD of the BMR model ................................. 32 3: Dimensional details of the rods .................................................................................... 38 4: Material properties of the rods...................................................................................... 38 5: Contact radius and area from the axisymmetric fine model in comparison to the analytical and experimental results................................................................................... 42 6: RMS errors of speed and strain of the axisymmetric FE model (error analysis performed against the analytical solution)........................................................................ 43 7: Material properties of the components of the BMR model .......................................... 54 LIST OF ABBREVIATIONS A/P - Anterior/Posterior BMR - Bio-Mechanical Reduction CAD - Computer Aided Design COC - Ceramic-on-Ceramic FEA - Finite Element Analysis ISO - International Organization for Standardization M/L - Medial-Lateral MOM - Metal-on-Metal MOP - Metal-on-Polyethylene NIH - National Instituted of Health PSG - Polar Separation Gap THA - Total Hip Replacement ACKNOWLEDGEMENTS First, I would like to thank my advisor, Dr. Rebecca Brannon, for her guidance throughout this project. Her persevering support both as my advisor and as a fellow woman has boosted my confidence. I would like to thank Dr. Dan Adams and Dr. Ken Monson for being a part of my supervisory committee and for their valuable suggestions. I would like to specially thank my mentor Tony Sanders for letting me be a part of the ‘hip' team and for his immense support and guidance throughout the life of the project. I would like to thank NIH for being the financial backbone for this research. I would like to thank Exponent, Inc. employees for their efforts in computational analysis. I would like to thank my colleague Ira Tibbitts for his help whenever I needed it and for investigating the calibration error and re-calibrating the drop tower apparatus. I would like to express my gratitude to Jeff Kessler for allowing us to use the machines in the Structural Integrity Laboratory at the University of Utah, and to the Ortho Development Corporation and the machinists for their experienced and high precision machining and for making this project relatively economical. I also would like to thank my close friends and roommates for their moral support. This section would be incomplete without mentioning the most important persons in my life - my parents and family members who confided in me and brought me up to this stage; Thank you Mom and Dad. Above all of us, I would like to thank the God Almighty for giving me the strength and courage to face all the challenges. CHAPTER 1 INTRODUCTION 1.1 Total Hip Arthroplasty Total hip arthroplasty (THA) is a surgical treatment that alleviates pain and disability by replacing a damaged, arthritic hip joint with a prosthetic ball-and-socket joint. The specific anatomical features that are surgically replaced are the femoral head and the acetabulum. A femoral stem implant is placed into a broached intramedullary canal in the femur's proximal metaphysis. A femoral head (the joint's ball member) is then assembled with the femoral stem implant. The acetabulum is replaced by a socket, typically assembled from two separate, nested hemispherical cups. The outer metal cup, termed the "shell," supports the inner cup, termed the "liner," which articulates with the head. The liner and the femoral head form the new bearing surfaces of the prosthetic hip joint as shown in Fig 1. 1.2 Implant Materials Hip implant bearings are typically made of plastics, metals, or ceramics paired together in various combinations. The plastic material is usually a modified form of ultra-high molecular weight polyethylene (UHMWPE), enhanced by radiation cross-linking (XLPE) to improve wear resistance. Metal bearings are made of cobalt-chromium (Co- Cr) alloy, which is hard, wear resistant, and biocompatible in bulk form. The most widely used ceramic materials in hip bearings are aluminum oxide (alumina) and zirconia- 2 toughened-alumina (ZTA), and (less commonly) zirconium oxide (zirconia). The ceramics are the hardest and the most wear-resistant of the standard implant bearing materials [2-4]. The various hip implant bearing couples are categorized based on specific combinations of all these materials; generally, the categories consist of Hard- Polyethylene (PE) and Hard-Hard couples, as shown in Table 1. 1.3 Background 1.3.1 COC Implants - Squeaking - Stripe Wear The ceramic materials used in hip prosthesis bearings offer properties giving them superior performance in some ways, yet they also have drawbacks that have prevented COC couples from becoming the preferred hip bearing option. Their principal advantage is their extreme hardness. Alumina bearings, for example, have a hardness exceeding 1800 GPa [5] which compares to 500 GPa or less for CoCr [6], the hardest metal hip bearings. Their high hardness is one property making COC couples the lowest-wearing category of hip bearing. In hip simulator wear tests, COC bearings wore at a rate of <0.01 mm3/Mcycles, MOM bearings at a rate of 0.1 mm3/Mcycles, and MOP bearings at a rate of 13 mm3/Mcycles [7]. The drawback of high hardness is brittleness, and there have been numerous reports of COC hips exhibiting sudden catastrophic fracture [4, 8- 10]. To combat such failures, ceramic implant manufacturers have improved the material's density and implemented overload proof testing of every unit [11]. Consequently, fracture failure rates have diminished to less than 0.01% according to one recent report [12]. Another as yet unresolved, shortcoming of COC hips is that they may become noisy after several months of in-vivo service; this problem is the subject of intense contemporary research. 3 There have been numerous recent clinical reports of various noises such as squeaking, clicking, or grating emanating from COC hips [13]. Among these noises, squeaking appears to be the most common and the chief source of patient complaints [13- 15]. In a recent review of squeaking in COC hips, 10.7% out of 131 patients reported audible squeak during daily activities [16]. Another recent clinical report has shown that 10.6% of the COC hips generated noise, out of which 2.6% were defined as squeaking [17]. Although squeaking has not been linked to painful or debilitating symptoms, the annoying noise leads to some affected patients to undergo a revision surgery in which, either or both of the bearings are replaced [16, 17]. According to Ranawat et al., squeaking is due to a combination of subluxation (a small-scale separation of the head from the liner, not great enough to result in complete dislocation), edge loading (contact between the head and the rim of the liner, in contrast with typical loading where the head contacts the liner's spherical inner surface), and subsequent stripe wear (a particular wear pattern of COC hips, further described below) [18]. According to Yang et al., the etiology of the squeaking problem is multifactorial, involving causes such as femoral head subluxation, femoral neck impingement on the acetabular liner, and secondary stripe wear [19]. Further research on the cause of squeaking shows that it could be a consequence of stripe wear [20] or else of particular design features of the implant system [13, 21]. Recently, metal transfer (caused by subluxation during which the femoral head slides onto the edge of a metallic shell), together with a disruption of fluid film lubrication, was reported as the sole repeatable cause of squeaking in a laboratory hip joint simulator [22]. However, ongoing (yet 4 unpublished) research at the University of Utah shows that subluxation and squeaking are not limited to metal components, but can also occur in COC bearing couples. Stripe wear is the term used to describe a particular distribution of wear seen on some retrieved COC bearing couples. Fig 2 illustrates a typical example of stripe wear on a femoral head; numerous clinical reports show that this type of wear, though typically shaped as thin and elongated, can manifest in a variety of shapes, sizes, and orientations [20, 23, 24]. Some investigators have reported that stripe wear occurs due to edge loading between the COC bearing surfaces [24]. Theories describing the cause of stripe wear are further discussed in Section 1.3.2. 1.3.2 Micro-separation / Micro-lateralization The typical human gait cycle may be subtly altered by THA in a way that could lead to the problems of COC stripe wear and squeaking. A walking gait cycle consists of two alternating phases for each leg: (1) stance phase, during which the leg supports the body via foot-ground contact, and (2) swing phase, during which the leg is swung forward to bear the next footfall. Researchers have imaged artificial hip joints in-vivo using fluoroscopic video examination of patients (Fig 3) performing a variety of common activities including walking [25-27]. They have detected that, in many activities, the prosthetic femoral head may temporarily sublux from the acetabular liner by a small distance, on the scale of fractions of a millimeter to a few millimeters. The occurrence of such separation during the swing phase of gait is explained as an effect of soft-tissue laxity and changes to the dimensions of the hip joint - both secondary to hip replacement surgery - as well as the non-load-bearing state of the leg during swing phase. Upon heel 5 strike at the beginning of stance phase, the subluxed head is forcibly relocated into the prosthetic cup; this motion is termed as "reduction" of the femoral head. Separate research work has incorporated this understanding of prosthetic hip kinematics in laboratory test methods, with the outcome of eliciting more clinically relevant wear patterns on tested hip implants. Traditionally, hip simulator studies used only fully reduced, concentric head-liner articulations, and the resulting bearing wear patterns did not match with in-vivo wear patterns [28]. In a novel experiment, Nevelos et al. implemented a design change to a hip joint simulator, inserting a spring to impart a medial-lateral spring force that caused supero-lateral subluxation of the femoral head relative to the acetabular liner during the low-load, swing phase of simulated gait [29]. Though the details of how and where the spring force had been applied are not reported in this article, we have adapted a similar mechanism of using a lateral spring force in our study to cause subluxation of the femoral head. The researchers termed the new motion "micro-separation", and by adding it to the hip simulator motion pattern, they succeeded in replicating clinically relevant in-vitro COC stripe wear [29]. Wear scars generated in-vitro via the addition of micro-separation motion have also been correlated with increases in gravimetrically measured wear [29-31]. Micro-separation is not limited to COC bearings; it is also a concern in hard-PE bearing couples [32]. There are multiple potential causes for the edge loading that leads to stripe wear in-vivo. One explanation is the swing-phase micro-separation is followed by forced edge loading contact between the bearing surfaces upon heel strike [20, 31]. Another explanation is that femoral neck impingement with the rim of the acetabular liner can cause bearing couple separation and edge loading [30]. Although the research in this 6 thesis focuses on edge loading at the beginning of heel strike, it should be noted that edge loading can also occur during other activities, such as standing up from a chair [24]. The term micro-separation was originally described and illustrated as an actual separation of the bearing surfaces [33]. Recently, authors from the same research center have published a more precise term "micro-lateralization," by which they mean to emphasize that the femoral head maintains contact with the edge of the acetabular liner throughout subluxation and edge loading [34]. Though numerous hip simulator studies have included micro-separation and achieved clinically relevant stripe wear, scientific understanding of the contact mechanics of prosthetic hip joints undergoing edge loading is still in a preliminary stage and, therefore, is a focus of this research effort. 1.3.3 Contact Force between Implants Reducing hip implant wear and averting problems such as squeaking require improved understanding of hip joint mechanics, including kinematics, kinetics, and contact mechanics. Edge loading is a forced contact between the ball and head that is not anticipated or safely accommodated by the bearing's design, and it occurs during daily activities such as normal human gait, sitting, stair climbing, walking, etc. [25]. Previous studies have shown that edge loading occurs with or without subluxation [35]. Edge loading can also occur as a result of steep cup angles associated with malpositioning and instability of the acetabular components [36]. We have used a steep cup angle (60°) in our study to represent a worst-case scenario of the cup angle that could cause edge loading. Average hip contact forces, correlated in time with gait patterns of daily activities, were examined in a few patients with prosthetic implants [37]. The average hip contact forces reported in this paper come from a study of a few patients and, hence, there 7 is a possibility that the average could vary drastically if more number of patients were studied. Contact mechanics between COC implants were studied under adverse conditions [33, 38]. In [33], only ideal conditions where the femoral head contacts within the acetabular liner were considered making it a static analysis. In [38], the contact stress distribution between COC bearing surfaces was analyzed using the finite element method, but that study failed to include a physical dynamic event (like the human gait cycle). The contact force between the bearing surfaces was coupled with micro-separation in causing implant wear in several hip simulator studies. A lower swing-phase load coupled with severe micro-separation yields a higher wear rate when compared to the wear rate caused by a higher swing-phase load with mild micro-separation [31]. However, a contact force between the bearing surfaces has not been quantified in all of the hip simulator studies involving the micro-separation and edge loading modes. Contact stress distributions were studied under static edge loading conditions and different micro-separation values using finite element analysis [33]. The degree of these contact stresses is highly dependent on various parameters such as the value of micro-separation, radial clearance between the bearing members, inclination angle of the acetabular cup, and thickness of the ceramic liner. Scrutiny of several of these studies reveals that there is presently a gap in scientific understanding about the mechanics of hip prosthesis edge loading, namely that there is little knowledge about the contact forces that accompany edge loading. Researchers who have implemented edge loading in hip simulator tests have done so by imposing a pre-selected micro-separation distance [29, 39]. Only recently has there appeared a published report that analytically determined contact forces in such tests [34]. Weak understanding 8 of edge loading mechanics, both in-vivo and in-vitro, impairs resolution of current problems such as COC hip squeaking. This shortcoming is rectified by the work in this thesis, which gives particular attention to forces associated with micro-separation. 1.4 Motivation The observed wear patterns on COC implants are important in two senses. First, they probably cause squeaking and hence the need for revision surgery, which is an acute problem. Moreover, these problems discourage widespread use of bearing materials that would otherwise benefit many patients by being longer lasting and more wear resistant. To overcome such problems, there is a need for better understanding of the severe, adverse load conditions, such as edge loading, that apparently cause the most severe, clinically observed wear. At present, there is little understanding of the forces that arise between the implants during the severe loading events, and that limits our ability to test designs and materials under worst-case conditions. Though a few research labs have implemented micro-separation conditions in hip simulators, such test methods are not yet widely adopted; for example, the ISO standard for hip simulator wear testing includes no description of or requirement for micro-separation conditions [1]. This lack of transfer from research to widespread practice may be due in part to a lack of understanding or acceptance of the micro-separation conditions, such as the separation distance or separation frequency, that those modified wear simulators impose. The publications about those simulators and their routines do not provide crucial design details, nor do they quantify the forcing functions [29, 34]; more fundamentally, the conditions imposed during these tests have not been demonstrated to be representative of in-vivo hip joint mechanics. Thus, there is a need for better understanding of the basic mechanics and 9 mechanisms involved in severe-condition hip testing, as well as the mechanics of adverse hip joint motions. The research literature shows that subluxation and consequent edge loading may occur under a wide variety of activities [25]. Each such activity could have its own conditions, and any individual patient could exacerbate those conditions via details such as bodyweight and, frequency of various activities. Yet, no design testing program can consider all possible scenarios, so it is vital that the testing which becomes standardized be representative of frequent events, such as the subluxation and reduction that happen during routine walking. One key missing element in our understanding of edge loading associated with micro-separation during walking is the contact force between the bearing surfaces. There is only one known report of an analysis that determines such a contact force, and that comes from a simulation of a test machine, not of a human body activity [34]. Having more information about the contact force during edge loading could result in more relevant test methods, and perhaps more widespread acceptance and use of those methods. Furthermore, it could open doors to innovative, physiologically relevant testing protocols that could accelerate progress in material and design development. To address the aforementioned gaps in COC implant wear data, we study a commonly occurring event that causes edge loading, to determine the contact force and the contact stress that arise during that event. We design a simple model to represent edge loading in conjunction with swing-phase subluxation and heel-strike reduction. In this model, a femoral head is initially subluxed and contacting the edge of the acetabular cup, as though at the end of swing phase just prior to heel strike. To emulate the conditions of 10 the gap geometry reported from in-vivo fluoroscopy [25], a small initial lateral force is applied to keep the head in contact with the edge of the cup. The head is suddenly relocated into the liner by an axially directed reduction motion, as would occur during the heel-strike event of the gait cycle. This biomechanical model of edge loading was examined using a combination of computational and experimental methods as further described throughout this thesis. 1.5 Objectives • The main objective of this project is to quantify the ensuing contact force between the bearing surfaces of COC hip implants. Further, the resulting contact stress distribution is analyzed. • Since the contact force cannot be directly measured from a physical laboratory version of the model, it is instead indirectly inferred using a finite-element based version that is validated against data that can be measured. Accordingly, a supporting project objective is to thoroughly validate the finite element model in a laboratory test equipped with sensitive dynamic diagnostics. • A further objective is to develop computationally cost effective techniques to accurately resolve contact stresses in the vicinity of relatively minute edge-loading contact patches. To this end, the FEA in this thesis employs an aggressive approach to submodeling the region directly surrounding the contact area between the head and the liner. • A preliminary objective was to acclimate the project's team of engineers to the computational and laboratory techniques for the biomechanical model study. 11 Hence, this thesis also reports an opening project to measure and analyze contact stresses and vibratory responses during the low speed impact of slender rods. Our hypothesis is that the contact force during the sudden reduction is strongly affected by the dynamics of stress wave propagation in the femur during the period of contact. This hypothesis is supported by observations of transverse and oblique impact of long, slender rods. Studies of such impact have shown that the peak contact forces are noticeably influenced by stress waves when the duration of contact is similar to the vibration period of the induced bending stress waves [40, 41]. We hypothesize that similar phenomena occur in a femur when a sudden, impulsive load is applied to its proximal end via a COC contact couple. This hypothesis comes in part because the femur is compliant in comparison to the implants, a property that favors relatively long vibration periods. It also comes from the fact that the COC contact couple is very stiff, a property which ensures that the contact displacements will be small and hence the contact duration will be very short. Since there are currently no methods for direct measurements of contact forces, finite element analysis is the primary means of investigating the validity of the key hypothesis. A simplistic means of estimating the contact forces would be to use a rigid body dynamics approach coupled with Hertzian contact theory. However, to account for dynamic stress waves, a more sophisticated approach is needed; so, a finite element model is analyzed using an explicit solution approach. A rigid body analysis is nevertheless developed to provide a comparison of the simplest relevant mechanical analysis with the results of the more sophisticated FEA approach. 12 1.6 Specific and Novel Contributions This thesis contributes to the body of knowledge in the fields of orthopedic implant design and testing, hip joint biomechanics, and COC hip bearing couples. The specific and novel contributions of this thesis include the following: • An engineering model of a common reduction event by which hip joint prostheses experience edge loading that may be a root cause of the currently observed problems of stripe wear and ceramic hip squeaking; specifically, the estimation of worst-case contact forces between the bearing surfaces. • Demonstration of the role played by dynamic stress wave propagation caused by impulsive loading events that affect COC hips and the skeletal members to which they are attached. It was revealed that the stress wave at the contact did not have a predominant effect on the contact force duration, as it was generated by an insufficiently rapid input given to the model. • A novel and aggressive approach to submodeling in finite element analysis of the stresses arising around very small contact areas during loading of hard-on-hard hip bearings. This approach, which has been demonstrated to be effective in other application areas [42], has not yet been widely practiced in the orthopedic research field. This work illustrated its merits by setting a new standard for the refinement that is both required and readily achievable for contact mechanics analysis in hard-on-hard orthopedic implants. Further improvement of the submodeling strategy is required to better quantify the contact mechanics in such problems involving contact. 13 • These results will help to better define the mechanical conditions by which hip joint wear studies can mimic worst-case (though common) in-vivo loading, which in turn is anticipated to contribute to improving the longevity of COC hip implants. Fig 1: Schematic of THA [43] Table 1: Available prosthetic hip material couples and their corresponding acronyms Head Material Liner Material Polyethylene Metal Ceramic Metal Metal-on-Polyethylene (MOP) Metal-on-Metal (MOM) - Ceramic Ceramic-on-Polyethylene (COP) Ceramic-on-Metal (COM) Ceramic-on-Ceramic (COC) Acetabular Liner Acetabular Shell Femoral Stem Femoral Head 14 Fig 2: Stripe wear (made visible by rubbing with graphite) on femoral head and acetabular liner from a retrieved COC hip [23] Fig 3: Radiographic image of an in-vivo hip with a superimposed CAD model in the position of the bearing couple (left); Schematic showing micro-separation of 4.3 mm (right) [26] • CHAPTER 2 BIOMECHANICAL REDUCTION (BMR) MODEL 2.1 Introduction This research has developed and used a simple model of a prosthetic hip joint and the immediately connected skeleton to study the mechanics of an edge loading event. The event was an impulsive reduction of a subluxed femoral head from an initial contact position on the edge of an acetabular liner. In its initial configuration, the model simulated conditions existing at the end of the swing phase of gait, just prior to heel strike. The reduction phase of the calculation predicts motion and forces occurring as an immediate reaction to heel strike. A key reason for choosing to model this particular edge-loading scenario is that it is the same scenario represented in micro-separation hip simulator wear tests developed at several different research centers [31, 39]. In research tests at those centers, the edge loading and reduction motions have been implemented for the purpose of studying implant wear under severe but realistic conditions in long-term, repetitive motion tribological tests. In contrast, the present study has developed a model of impulsive reduction for the purpose of studying the instantaneous mechanics of the event, i.e. the effects occurring within several milliseconds. The engineering model used to represent this event is herein referred to as the Bio-Mechanical Reduction (BMR) model. 16 This chapter describes and illustrates the BMR model in its abstract sense as a simplified representation of a complete physical entity, the simplification comprising only those elements that substantially influence the complete entity's response to a specified mechanical stimulus. Later chapters describe methods to implement the model in both experimental and computational contexts. 2.2 Components of the BMR Model The BMR model is an abstract representation of a human pelvis and a femur that undergo the sudden reduction event. The model consists of two subassemblies - a femoral subassembly and a pelvic subassembly. The femoral subassembly consists of a synthetic left femur implanted with a press-fit femoral stem and a femoral head. The pelvic subassembly consists of a rigid polyurethane foam test block to represent the pelvis; the block is prepared with a hemispherical hole representing the acetabulum, and the hole is implanted with a press-fit acetabular shell and a corresponding acetabular liner. Further details describing the actual implants employed in this model are given in the chapters describing the experiments. The following sections in this chapter describe the model's inputs and constraints. Though the muscles and tendons in the human body play the role of keeping the skeleton intact and generating voluntary skeletal movements, an underlying assumption of the BMR model is that the forces and constraints imposed by the soft tissues may be neglected in comparison to the forces between hard components (bone and implant materials) for the purposes of the event under study. There are three key points that support this simplification in the model. 17 • The reduction event occurs as a reaction to foot-to-ground contact (heel strike); this force gets transferred in a very short span of time from the heel to the hip joint via the tibia, the knee joint and the femur. In that short span, the reaction forces acting to change the subluxation state at the hip joint are the forces traveling through the skeletal members and not the forces of voluntary muscle contractions. The BMR model models the skeleton-propagating reaction force and neglects the soft tissue reactions. • The event under study lasts for a few milliseconds. It is hypothesized that during this short span of time, the contact force will be affected by a few reflected stress waves in the surrounding components. The stress waves in the stiffer components (skeletal members), which travel at higher velocity than those in the soft tissues, will have a dominating effect on the contact force. Moreover, stress wave attenuation caused by viscous dissipation in the soft tissues can be neglected because the duration of the event is considerably smaller than the characteristic response time of the material. In other words, the Deborah number (i.e., the ratio of the material's relaxation time to the event stimulus time) is large thus, justifying an approximation that the soft tissues respond elastically. • In an ideal THA, the muscles and ligaments around the hip joint help to keep the femoral head (ball) and the acetabular liner (socket) in a fully reduced articulation. However, some THA surgeries may result in a relatively loose joint due to muscle and ligament weakness or laxity, allowing subluxation of the head from the liner. When the subluxation occurs, the muscles take up the laxity by exerting force on the hip joint (similar to the force exerted by muscles 18 surrounding a normal hip) and cause a contact between the head and the liner's edge. During the heel-strike phase, the contact force generated between the femoral head and the liner's edge rises so high that it cannot be attributed to the small magnitude voluntary contractions of the muscles surrounding the hip joint [44]; therefore the BMR model does not implement separate actuators for such muscle forces. 2.3 BMR Model Configuration The BMR model, as shown in Fig 4, was configured to mimic certain conditions specified within an international standard for hip simulator wear tests, ISO 14242-1 [1]. Specifically, that standard describes a cycle of synchronized motions and loads at the hip joint that emulates the kinetics of the hip during normal walking gait. The characteristics of that cycle that were adopted into the BMR model were the relative orientations of the femoral stem and acetabular liner at the beginning of the gait cycle as shown in Fig 5. A macro view of the configuration of the BMR model, with the femoral head fully reduced in the liner, is given in Fig 6. This configuration was achieved by applying a specific sequence of rotations to the pelvic and the femoral subassemblies relative to the reference coordinate frame and neutral orientations described by the standard. That sequence is described in brief below and in detail in Appendix A. Furthermore, the 3D BMR model was designed to permit only 2D planar (X-Y) rigid-body motions, with the motion plane being a coronal anatomic plane through the center of the fully reduced head-liner couple. Out-of-plane motions were neglected for simplicity. Section 2.5 provides rationale for these aspects of the model design. 19 2.3.1 Standard Orientations This section describes the procedures followed to place the model into the initial configuration that is consistent with the ISO standard, with the exception that the abduction angle was greater than that described by the standard. First, the acetabular liner was rotated to an abduction angle of 60°, which was 15° beyond the conventional 45° angle specified in the standard; the additional 15° was implemented to model a severe but realistic liner orientation that increases the propensity for subluxation and edge loading. Next, the acetabular liner was rotated by an anteversion angle of 20°. Then, the femur subassembly was oriented in the following steps: 3° of adduction, then 4° of net external rotation, and finally 25° of flexion. 2.3.2 Feasible Setup Orientations The procedures described in the previous section produced a configuration in which the load line (represented by the line labeled "L" in Fig 5a) is at a certain angle from the vertical. This section provides the motivation and the procedures to reorient the configuration to best suit the constraints of the testing equipment. The full assembly was rotated such that the load line was vertical, and it was inverted because (1) the test frame is set up such that it applies vertical loads, (2) to be stable under self-weight, the femur is suspended rather than supporting its weight from the bottom and, (3) the actuator is bottom-mounted in the available test frame. These orientations include rotating the full assembly by 20° about a medial-lateral (M/L) axis, followed by orienting the load line to a vertical orientation as illustrated in steps 7 and 8 in Appendix A. Finally, the BMR model assembly was anatomically inverted, because this orientation better suited the configuration of the test machine and the fixtures described in later chapters. 20 2.3.3 Final Configuration The BMR model as configured for experimental trials, after applying the orientations previously described, is illustrated in Fig 6. The pelvic subassembly was angled at 40° with respect to an X-Z plane (a transverse anatomic plane). The required femur orientation of 8.6° was implemented in the experiments using a fixture that provided the 64 mm horizontal offset between the distal pin axis and the vertical load line. The load line passed through the center of the acetabular liner. This configuration was implemented in the FEA using software inputs of the dimension values. 2.4 Initial Conditions The polar separation gap (PSG) is the distance between the liner's internal spherical surface (its ID) and the head, measured parallel to the liner's axis of symmetry, as shown in Fig 7. The means of implementing the desired PSG dimensions are described in Section 4.8.2 for the experiments and in Section 5.4 for the FEA. 2.5 Inputs and Boundary Conditions The model's inputs and BCs were designed to cause and constrain rapid reduction of a subluxed femoral head. In the human body, the pelvis, the subluxed hip joint, and the knee joint at the distal end of the femur each have six degrees of freedom (DOF) and are acted upon by numerous restraints (e.g., ligaments and opposing joint surfaces) and active elements (various muscle groups). Creating a high-fidelity model representing all such BCs would be a tremendous undertaking beyond the scope and resources of this project. However, because the reduction event involves a short (roughly 2 mm) femoral head trajectory with only small displacements and rotations of the pelvis and femur, it 21 was feasible to faithfully model the reduction motion by implementing limited degrees of freedom in the model's BCs. Furthermore, the sudden femoral head reduction is essentially a reaction to external forces rather than a motion instigated by muscular contractions. It is a reaction to gravity acting vertically on the body and the ground reaction force acting vertically through the femur via the foot and the tibia at the instant of heel strike. Therefore, the BMR model simplified the BCs to act along a single, vertical axis, and it neglected any input from voluntary muscle contractions. The model's third key BC was a passive element (lateral spring force, Fig 4) that instigated the initial femoral head subluxation along with a small edge-load contact force. Each of these three key BCs is further described below, and their specific realizations in the experimental and analytical contexts are described in later chapters. In-vivo, the distal condyles of the femur (Fig 8) are seated on the tibial condyles. During normal gait, when the heel strikes the ground, the reaction emanating from the heel strike gets transferred through the femur to the hip joint and causes the femoral head to relocate into the acetabular liner. During this reduction motion, the femoral head moves only a short distance relative to the pelvis. In the BMR model, this motion is effectively represented by rotation about an axis allowing a single rigid-body DOF of the femoral subassembly. Hence, the distal end of the femur is fixed except for rotation about an axis along the anterior-posterior (A/P) direction passing through the distal metaphysis and between the condyles. This distal pin axis is shown in Fig 8. The muscles and tendons around the skeleton in-vivo constrain the movement of the skeleton. The soft tissues surrounding the hip help the femoral head in constraining its 22 movement within the acetabular liner. In the BMR model, a lateral force is used to maintain contact between the femoral head and the liner during subluxation. Specifically, this force is applied horizontally in the frontal plane at an anatomical position below the femur's greater trochanter (Fig 9). In-vivo, the acetabulum is supported by the pelvis. In the BMR model, the acetabular shell is press-fit into a polyurethane block representing the pelvis (Fig 10). All the components of the pelvic subassembly move together, as controlled by a dynamic input provided by an MTS actuator. During the heel-strike phase, the contact force between the bearing surfaces is a result of the ground reaction force at the heel. The input given to the acetabular subassembly in the BMR model causes a forced reduction of the femoral head into the liner, similar to the reduction caused by heel strike in normal human gait. 2.6 Rigid Body Analysis of BMR Model Rigid body analysis is a valid and applicable analysis approach when material deformations are insignificant. The BMR model involves biological materials which are compliant in nature, and the material deformations might have a substantial effect on the system response. Nonetheless, a rigid body analysis of the BMR model was performed to provide a low-order estimate of the model's response to the applied forces and moments. Specifically, the effect of the lateral spring force on the contact force between the bearing members was estimated in this analysis. This analysis showed us that, with the present design of the BMR model, the effect of the spring force on the contact force is greater than the effect of the inertia of the femoral subassembly on the contact force. 23 2.6.1 Equations of Motion Fig 11 shows a free body diagram (FBD) of the BMR model illustrating the applied forces and reaction forces. Table 2 describes the symbols used in the FBD. Since this analysis is limited to planar rigid body motion, the femoral subassembly is considered to be rotating about point ‘O' (the distal pin axis described in Section 2.5). The components of the model other than the directly contacting members (femoral head and acetabular liner) were represented using simple geometries. Accordingly, a line connecting the distal pin axis and the femoral head center represented the femur and the femoral stem geometries. In the actual experiment, the contact force acting on the femoral head was generated by controlling the test machine's actuator to provide a pre-defined dynamic velocity record. According to Newton's Second Law, Σπ ππ − ππ = 0 (1) where Fi are the applied forces, m is the mass of the femoral subassembly, and a is the acceleration of the center of mass of the femoral subassembly. Recognizing that sliding is occurring, Newton's second law, applied in X and Y directions, gives respectively πΆπ(μ cos θ − sin θ) − π
π₯ + πΆπ − πππ₯ = 0 (2) πΆπ(cos θ + μ sin θ) − π
π¦ − π − πππ¦ = 0 (3) According to D'Alembert's principle, the moments about point O can be written as Σ ππ − πΌπ πΌ = 0 (4) where Σ ππ is the sum of the moments of all the forces (acting on the system) about point O, Io is the mass moment of inertia about point O, and α is the angular acceleration. 24 For the simple dynamic analysis, Eq. 4 can be written in terms of the forces and the moment arm lengths (labeled in Fig 11) as (ππ‘ − π) πΆπ + π πΆπ − π₯ π − πΌππΌ = 0 (5) Eq. 5 was further analyzed by substituting the appropriate values of the applied forces and the corresponding moment arm lengths to determine the effect of the spring force on the contact force. 2.6.2 Effect of the Spring Force on the Contact Force Equation 6, (manipulated version of Eq. 5), was analyzed to estimate the effect of the spring force on the contact force. πΆπ = π (π−ππ‘) πΆπ − π₯ (π−ππ‘) π + πΌπ (π−ππ‘) πΌ (6) The objective of the rigid body analysis was to estimate a contact force value for a given value of the spring force and to further estimate how the contact force varied over a range of spring force values in order to better understand the system response. The spring force (CS) value was treated as a constant, which neglected changes due to extension of the spring (+2 N variation) during the reduction event. The mass properties (W and Io) of the femoral subassembly were determined from its 3D model in SolidWorks (Dassault Systemes, SolidWorks Corp., Waltham, MA). The angular acceleration of the femoral subassembly was computed based on a linear acceleration at the contact point. The linear acceleration value was determined using the derivative of the velocity recorded in an experimental trial with the BMR experiment configuration. The coefficient of friction (μ) was 0.06, a value taken from the literature [45]. These values were substituted in Eq. 6 to estimate the contact force (CN). 25 The rigid body model was also used to qualitatively analyze the effect of coefficient of friction (μ) on the velocity of the femoral head as described in the Discussion chapter. Equation 6 was analyzed by substituting the corresponding values of the inputs (on the right hand side of Eq. 6) and a range of spring force values (0 - 200 N). Though the rigid body analysis performed was a simpler version of the actual model and the estimations in this analysis are not comparable to the nonrigid analysis, it was concluded that the BMR model was sensitive to the value of the spring force chosen. As seen in Fig 12 the contact force increased as the spring force increased. Nevelos et al. [29] used a lateral spring force of 400 N (though the details of how and where the spring force actually was applied were not mentioned). The same value of the spring force in the rigid body analysis of the BMR model gave a contact force estimate of about 800 N. The estimation of contact force using the rigid body analysis provided us the background to design the lateral spring, i.e., the spring force value, used in the BMR model. Micro-separation was achieved using a relatively low value of the spring force, 105 N; this value was about 1/4th of a similar lateral spring force used in hip simulators previously [29]. There are no springs in human bodies, but a spring in the BMR model was incorporated to achieve the desired PSG and to maintain a forced contact on the rim of the liner. Also, even though the femoral head subluxes, the femoral head and the acetabular liner rim are in forced contact (as determined by fluoroscopic imaging [26]) due to the low magnitude forces exerted by the soft tissues. The lateral spring, though not physically present in the human body, represents the net effect of the soft tissue forces in 26 high Deborah number loading, and it maintains a forced contact between the head and the liner during subluxation, to mimic clinically observed prosthetic hip joint kinematics. The BMR model includes biological components, such as the femur, which are more compliant than the implants. So, we think that the femur's compliance will have a significant effect on the contact mechanics, which is the key hypothesis. Hence, there is a need to include the compliance in the components of the model. So, the BMR model was examined mathematically and experimentally as described in further chapters. X Y Fig 4: Illustration of the BMR model 27 Z X Z Y Y X Y X Fig 5: ISO 14242-1 standard - followed to orient the BMR model [1] 28 Y X Y Z Fig 6: Macro view of the BMR model Fig 7: The polar separation gap (PSG) 29 Fig 9: Lateral force application to achieve a forced contact between the femoral head and the acetabular liner Lateral force Y X distal pin axis distal condyles Fig 8: A/P axis of rotation in BMR model 30 Fig 10: The pelvic subassembly showing its components - acetabular liner, acetabular shell, and model pelvis (rigid polyurethane test block) Acetabular Shell Test Block Acetabular Liner 31 M x O l Rx CN CT X Y CS Ry η θ s t n r Load line Femoral head Liner W Fig 11: Free body diagram of the BMR model (liner partly shown) 32 Table 2: Description of the symbols used in the FBD of the BMR model Symbol Description of the symbol O Distal pin axis CN Normal contact force CS Tangential contact force (due to friction) CT Applied lateral spring force Io Inertia of the femoral subassembly about point O Mo Moment about the pivot point Rx, Ry Reaction forces at the pivot point in X and Y directions respectively W Weight of the femoral subassembly l Length of the liner joining the distal pin axis and the center of the femoral head n Length of the moment arm of CN r Radius of the femoral head s Length of the moment arm of CS t Length of the moment arm of CT x Length of the moment arm of W μ Coefficient of friction η Angle subtended by CN with the vertical θ Angle subtended by the line joining the distal pin axis and the center of the femoral head with the vertical Fig 12: Contact force vs. Spring force according to the rigid-body analysis; for reference, the data point on the plot shows the contact force for the spring force value used in the BMR test 0 20 40 60 80 100 120 140 160 180 200 0 50 100 150 200 250 300 350 400 Spring Force (N) Contact Force (N) X: 105 Y: 209 CHAPTER 3 ELEMENTARY ANALYSIS AND TESTING1 3.1 Introduction A low-speed structural axial impact test of two slender rods was conducted in a drop tower to validate corresponding models of contact mechanics. Analytical and finite element (FE) models of the contact and vibration were validated by the laboratory testing results. The following describes the purpose, scope, method of testing and some preliminary analyses carried out before the test. The purpose of this test was to examine the contact-coupled longitudinal impact mechanics between two rods. From the predictions of this simple test and the adaptability of the diagnostics, this test serves the purpose of substantiating instrument calibration and establishing a strong theoretical and laboratory framework for using the same methods in an experimental and computational study of a more complicated biomechanical analysis (detailed in later chapters). It was hypothesized in this elementary testing that the contact duration of the rods would be twice the transit time of the stress wave initiated at the contact. A similar effect was expected in the biomechanical analysis of the femur. 1 Much of the material in this chapter is repeated from [44]. 34 3.2 Problem Statement As illustrated in Fig 13, a first rod, Rod 1, centrally impacts an initially stationary second rod, Rod 2. Both the rods are placed coaxial to each other. Rod 1 is spherical at its impacted end and has a flat distal end. Rod 2 has both ends flat. Dimensional details are given in Table 3 and Table 4. Impact of Rod 1 onto Rod 2 gives rise to a small contact area. Away from the contact area, all surfaces are traction free. To prevent rotation of the rods in the experiment, the distal ends are outfitted with o-rings or tape, which impose negligible lateral stress in comparison to the axial stress. The objectives of this test were to measure (1) the contact area, (2) the wave propagation speed and, (3) the strains at the midpoint of Rod 2. 3.3 Methods The problem was solved using finite element analysis (FEA), which was verified against an analytical solution and was validated against laboratory testing. 3.3.1 Laboratory Testing A schematic of the experimental setup is shown in Fig 14. A total of four rods (three of Rod 1 and one of Rod 2) were used to carry out the test with repeated trials. Both rods were made of precision ground A2 tool steel, hardened and tempered to Rc 60, from a single lot. The isentropic elastic material properties of the rods were determined from testing the rods by the impulse excitation method, ASTM E 1876, using a Grindosonic MK5 instrument (Lemmens, Lueven, BLG). The impact test was conducted in an Instron drop tower apparatus (Model 8250 HV, Instron Corp., Norwood, MA). The apparatus consisted of a motorized latch block from which a sled is suspended. As per the computer command, the sled may be released from 35 the latch block into free fall sliding along two columns. Rod 1 and Rod 2, both shown in Fig 15, were suspended in tubular fixtures as shown in Fig 16; these fixtures mounted on the sled and the base of the test apparatus. Both fixtures had o-rings on the inside diameter (shown in detail A of Fig 16) that gripped each rod just tightly enough to hold each static and had sufficient clearance behind the rods' distal ends for the rods to travel freely after the impact. The weight of Rod 1 was suspended by a thin strip of tape wound around its distal end to prevent it from sliding out of the fixture; otherwise, Rod 1 was distally unconstrained. Rod 2 was distally placed on a plastic plug press-fit into the tube. Upon impact, the friction holding the rods in place by the o-rings was overcome, and the rods each displaced axially. The quantities measured in the laboratory testing were the contact area on Rod 1 and the wave propagation speed and the strain in Rod 2. The measurement techniques are described below. A 3D laser Doppler vibrometer (CLV-3D, Polytec, Germany) was used to measure the velocity by focusing the laser tip on the mid-point of Rod 2, as shown in Fig 17a. Two foil strain gauges installed at the midpoint of Rod 2 (as shown in Fig 17b) were used to measure the strain. A strain gauge signal conditioner (Model 2310B Vishay Measurements Group) amplified the strain values obtained in the test. A data acquisition system (DAQ, USB 1604HS, Measurement Computing, Norwood, MA) interfaced with LabVIEW (National Instruments, Austin, TX) was used to acquire data. The rods were wired with 24g wire into a 3V electrical circuit that was also wired into one channel on the DAQ. Rod impact completed the circuit and triggered the start of data collection. 36 3.3.2 Axisymmetric Finite Element Modeling A 2D axisymmetric model of the rods and the 2D meshes was generated using LS-Prepost (Livermore Software Technology Corp., Livermore, CA) and consisted of 2D 4- node solid elements. Contact between the rods was modeled as a 2D surface-to-surface contact. The FEA had three mesh refinements of Rod 2 to examine convergence. The element aspect ratios were approximately 1:1, and the meshes were refined by serially halving the average length: from 0.42 mm (coarse), to 0.21 mm (medium), to 0.105 mm (fine). The finite element code LS-Dyna (Livermore Software Technology Corp., Livermore, CA) was used to solve the axisymmetric model and its refinements. Post-processing was done using LS-Prepost. 3.3.3 3D Finite Element Analysis2 To establish confidence in the methods used to analyze the 3D femur, the two-rod impact test was also analyzed using a 3D model [46]. This model used hexahedral elements, penalty contact algorithm and explicit time integration. A submodel taken from the vicinity of the contact region in the initial global model was used to further refine the analysis. Three mesh refinements of the submodel were used to assess convergence. This model also was verified using an analytical model and validated using the above mentioned experiment. The accuracy of the results in this simple analysis using the submodeling technique established confidence that a similar technique could be used in the computational study of the complicated impact mechanics of artificial hip joints. 2 Performed by Exponent, Inc., under our direction. 37 3.4 Results: Comparison of Analytical, Experimental and Axisymmetric FEA The results compared are (1) speed at the midpoint of Rod 2, (2) strain at the midpoint of Rod 2 and, (3) contact area on the rounded surface of Rod 1. Results from two experimental trials each of Rod 1 specimen are summarized in [46]. The experimental results provided here are those of Rod 1, Specimen 1, Trial # 3. The contact area on the rounded surface of Rod 1 in the experiment, shown in Fig 18 a and b, was recorded by a fingerprinting technique [46]. In the axisymmetric model, the distance between two extreme nodes of the line of maximum stress at the contact was measured using LS-Prepost. The contact area on Rod 1 is from a single mesh generated fine enough such that three elements of the slave (Rod 2) are in contact within an element of the master (Rod 1). Table 5 compares the contact radius and contact area measured by the three methods. The speed and the strain data both recorded at the midpoint of Rod 2 are overlaid on the analytical [46] and axisymmetric FEA results, as shown in Fig 19 and Fig 20, respectively. Convergence was studied on the 2D axisymmetric model for the recorded speed and strain. The errors in the computational and the experimental methods were determined against the analytical solution by taking a root mean square (RMS) difference of the quantities over time, with results tabulated in Table 6. The values show a decrease in the RMS error with mesh refinement. The difference in the RMS errors for the strain is not uniform through the mesh refinement because limitations of the software's data output capability required interpolating the FEA time histories. 38 3.5 Conclusion The results of the longitudinal impact of slender rods analyzed using the three methods (analytical, experimental, and FE) gave high reliability to out methods for analyzing and measuring the dynamic structural response of a simple system. This test performed in the laboratory served as a major platform to faithfully apply the same laboratory techniques in studying the transient dynamics of prosthetic hip joints. For FEA involving intricate shapes like biological components (for example, the femur), solving a 3D dynamic problem involving millions of elements and nodes might be computationally inefficient. Also, in contact problems, contact mechanics cannot be directly measured. For example, contact force cannot be directly measured experimentally. As in [46], the submodeling FE approach taken to solve this simple problem of two-rod impact - with computational efficiency - may be followed in solving complex problems such as those involving artificial hip joints. Table 3: Dimensional details of the rods Material Length (mm) Diameter (mm) Tip radius (mm) Impact speed (m/s) Rod 1 A2 tool steel, Rc 60 250.99 (l1) 12.7 (d) 35 (r) 2.197 (u1) Rod 2 A2 tool steel, Rc 60 700.99 (l2) 12.7 (d) flat Initially stationary (u2 = 0) Table 4: Material properties of the rods Young's modulus Poisson's ratio Density 204.3 GPa 0.3 7803 kg/m3 Rod 1 d u2 = 0 d u1 r Rod 2 l1 l2 Fig 13: Schematic of the impacted rods 39 Fig 15: Rod 1 (shorter) and Rod 2 (longer) used in the impact test; a US quarter is shown for reference to set scale trigger latch block sled columns Rod 1 Rod 2 Rod2 fixture strain gages drop height Rod 1 fixture Laser vibrometer clearance 24 g wire Fig 14: Schematic of the two rod impact test 40 Fig 16: Fixtures to hold Rod 1(top) and Rod 2 (bottom) 41 a) b) c) c) Fig 18: Contact patch area on the round surface of Rod 1 during the impact, recorded a) from a hand-held digital camera, b) from optical CMM; c) Stress contour plot from which the contact radius was measured is shown for the fine mesh of 2D axisymmetric model of Rod 1 and Rod 2 with their axes pointing downward (full length of the mesh not shown); Black horizontal lines on a corner of each figure shows scale to 1 mm. Fig 17: Laser focused at the midpoint of Rod 2 (left); Strain gage installed at the midpoint of Rod 2; second strain gage installed on the opposite side of the first one (right) 42 Table 5: Contact radius and area from the axisymmetric fine model in comparison to the analytical and experimental results Analytical FEA Experimental Δ, Analytical vs. FEA / Experimental Contact radius (mm) 1.09 1.1 1.109 -0.9% / -1.7% Contact area (mm2) 3.73 3.799 3.86 -1.8% / -3.5% Fig 19: Axisymmetric fine mesh results of speed measured at the midpoint of Rod 2 overlaid with analytical and experimental data 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 -0.2 0 0.2 0.4 0.6 0.8 1 1.2 Time (ms) Speed (m/s) Analytical FEA Exp't 43 Fig 20: Axisymmetric fine mesh results of strain measured at the midpoint of Rod 2 overlaid with analytical and experimental data Table 6: RMS errors of speed and strain of the axisymmetric FE model (error analysis performed against the analytical solution) Axisymmetric FE model RMS Δ speed (m/s) RMS Δ strain (με) Coarse 0.3 78.68 Medium 0.285 78.12 Fine 0.27 60.82 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 200 100 0 100 200 Time (ms) Strain (με) FEA Anaytical Exp't CHAPTER 4 LABORATORY TESTING OF BMR MODEL 4.1 Introduction A physical rendition of the BMR model was tested dynamically in the laboratory to measure its response to sudden femoral head reduction. Three separate response parameters were measured: (1) the velocity of the femoral head, (2) the strain on the femoral neck, and (3) the strain on the femoral shaft. The experimental setup is shown in Fig 21. The model was tested using a servohydraulic test frame (MTS, Eden Prairie, MN). Strain gages measured the two strain parameters, and a laser Doppler vibrometer measured the femoral head velocity. Trials were performed for three different PSG values: 2, 3, and 4 mm. The measured response parameters would later serve to validate a finite element simulation of the BMR model undergoing the same sudden reduction event. 4.2 The Femur Construct Researchers have substituted composite femurs for cadaveric ones because the composite femur's consistent properties and shape are conducive to repeatable biomechanical testing and analyses [47, 48]. Fourth generation composite femurs (4GCFs, the most recent design) are commercially available (Pacific Research Labs, Inc., Vashon, WA) both as physical models and as 3D CAD (computer aided design) models. Researchers prefer composite femurs over cadaveric femurs due to various reasons - 45 (1) composite femurs have very little interspecimen variability, especially compared with natural femurs, which vary greatly from person to person, (2) composite femurs are easily available and, (3) composite femurs are easy to store. Since the introduction of composite femurs in 1987, research scientists have improved the composite femurs from the 1st generation to the 4th generation. The cortex of the 4GCF is made up of short glass fiber reinforced epoxy. The material properties such as stiffness, tensile strength, and fracture toughness of the 4GCF cortex were tested and reported to be close to human femur cortex material [49]. One such commercially available 4GCF was used throughout this project. A femur construct (femoral subassembly) was prepared by implanting the composite femur (Model 3406 large left) with prosthetic hip implants. The femur was prepared using mock surgical techniques following the directions of the femoral stem manufacturer. The femoral head was removed with an oscillating saw, the distal intramedullary canal was cylindrically reamed, and the proximal metaphysis was broached with successively larger broaches. The proximal portion of the greater trochanter was removed to provide a line of sight to the laser used for femoral head velocity measurements. The trochanter is not normally resected in THA; however, its removal was a reasonable modification of typical surgical technique, because the trochanter lies outside of the load-bearing path when the femur is axially compressed. A proximally porous coated, titanium femoral stem (EncompassTM, Ortho Development, Draper, UT) was press-fit into the surgically prepared femur. The particular size of femoral stem was chosen based on the sizes of the final distal reamer and final broach used in the surgical preparation. A Ø36 mm Al2O3 femoral head (Biolox Forte, Ceramtec, Germany) was mated with the femoral stem trunnion. The proximal end of the femur 46 construct is shown in Fig 22. A dowel pin press-fit into the distal end of the femur between the condyles served as the distal pin axis described in Section 2.5. The material properties of the synthetic femur, the prosthetic hip implants, and the test block used to hold the acetabular components are summarized in Table 7. Note that the femoral cortex properties listed in Table 7 are manufacturer-provided values that were not used in the FEA analyses; more accurate values were determined by a calibration process described in Chapter 5, Section 5.3. 4.3 Fixtures and Initial Alignment Custom-made fixtures suspended the femur construct by its distal end from the test frame's load cell (refer to labels in Fig 21). The adapter plate provided the 64 mm horizontal offset, labeled in Fig 21, needed to configure the femur per the ISO standard initial conditions. The U-bracket held the femur by means of the press-fit dowel pin shown in Fig 23. The adapter plate and u-bracket were squared and tightened to each other prior to assembly in the test frame. They were tightened to the load cell after first rotating the entire femur and fixture subassembly to align the femoral stem's coronal plane with that of the acetabular liner, as visually judged. Spherically tipped set screws were threaded through the arms of the U-bracket and mated with the ends of the femur's distal pin, providing greased joints and forming the femur's rotation axis. At the bottom of the entire assembly, a thick aluminum base plate was attached to the top of the load frame's actuator. A sine plate was clamped to the base plate using strap clamps. The sine plate was angled at 40° to set the required acetabular abduction angle. Furthermore, the sine plate was aligned on the base plate such that the centerline of the actuator passed through the center of the acetabular liner. This alignment was established 47 using known dimensions of the base plate and precision gauge blocks that were contacted against a femoral head sitting in the liner. To set the anterior-posterior (A/P) position of the liner relative to the femur, the pelvic subassembly was positioned on the sine plate, pushing it with a micro-threaded adjustment screw (attached to the sine plate) until a trial edge contact, between the femoral head and the liner's edge, was observed to occur on the lowest part of the edge. This alignment was judged visually from the liner's lateral side and was aided by illuminating the interior of the liner with a strong pen light from the medial side of the femoral head. This careful alignment was necessary to ensure that the femoral head was centered with respect to the liner and would slide along the liner's coronal plane during the reduction tests, as it would in the finite element simulation. Once this alignment was achieved, the pelvic subassembly was fixed to the sine plate using strap clamps, which were added to remove visible slipping that was (without them) evident when using only the locking mechanism of the sine plate itself. The fixtures' stiffness was determined in a separate static test. The BMR model was entirely assembled as if for an actual trial. The test frame was rigged with a dial indicator (resolution 0.0001 in.) on an articulating arm. The indicator was posed horizontally and its tip was placed in contact with the femur's distal pin axis (i.e., one set screw protruding from an arm of the U-bracket). The actuator was displaced until the load cell registered a compressive load approximately equal to that observed in an actual reduction trial, and the deflection of the pin axis was measured from the dial indicator. A static free-body-diagram analysis of the fixtures revealed the horizontal force on the pin, and this value, divided by the measured horizontal pin deflection, gave an effective horizontal stiffness of the fixtures. A similar procedure yielded the fixtures' effective vertical stiffness. These 48 values were later implemented to effect a realistic compliance at the distal pin axis in the FEA of the BMR model. 4.4 Spring Mechanism A spring mechanism (refer to labels in Fig 21) was used to establish a lightly loaded contact between the femoral head and the edge of the liner, as illustrated in Fig 9. The connecting rod (Fig 24) ran through the inside of the compression spring, and the end of the rod was hooked together with the hook that was screwed into the lateral side of the proximal femur. The compression spring was squeezed between the adjustment nut and the rotation block (labeled in Fig 24a) through which the connecting rod passed, and this mechanism applied a laterally directed force to the femur, thus pulling the head into contact with the liner's rim. The mechanism permitted the connecting rod to slide in the rotation block, and the rotation block could rotate about a fixed axis. Thus, the connecting rod could slide and rotate to follow the movement of the proximal femur without constraining the femur's movement, other than via the spring force. The mechanism included a donut load cell (LC8150-500, Omega, Stamford, CT) to measure the force, along with an amplifier and a voltmeter to register the load cell's output. The initial spring force - established in the subluxed, edge-loaded condition - was 105 N in all trials. 4.5 Laser Vibrometer The velocity of the femoral head during reduction into the acetabular liner was measured using a laser Doppler vibrometer (CLV-3D, Polytec, Germany). The laser head of the vibrometer was supported on a tripod equipped with a 3D rotation head. The laser was focused on a small piece of retro-reflective tape stuck on the femoral head as shown 49 in Fig 25. The focus was adjusted by manually moving the tripod until the vibrometer's three lasers converged at a single spot, approximately Ø0.5 mm. Although the vibrometer could measure three orthogonal components of velocity at the focus point, only the component parallel to the laser head (the Z component per the vibrometer's terminology), was recorded, since that component was by far the greatest, as effected by the experimental design. The vibrometer provided a dynamic voltage output that was later scaled to a velocity signal using the system's calibration factor. 4.6 Strain Gaging Two foil strain gages were used to measure dynamic strain response in the femur construct during the sudden reduction tests. One was installed on the medial femoral diaphysis; this location was selected for maximal signal strength, since the location would experience maximal strains from combined compressive axial and bending loads. The other gage was installed on the medial side of the femoral neck; this location would also experience combined axial and bending loads, though lower strain magnitudes due to the stiffer substrate (titanium stem). The installed strain gages are pictured in Fig 26. For each gage, a corresponding dummy gage was installed on a dummy specimen of the same material to compensate for temperature effects when the gages were connected into a Wheatstone bridge circuit. High resistance 1000 Ω gages were selected since the synthetic femur was a poor heat-sink material. A high bandwidth signal conditioning amplifier (2310B, Vishay Micro-Measurements, Raleigh, NC) completed the bridge circuits. The amplifier was used with relatively low gain settings (e.g. <1000) to maximize the bandwidth and minimize potential attenuation of rapidly changing dynamic strain signals. The amplifier settings were calibrated using on-board shunt calibration. 50 The strain measurements in the two locations were made in separate trials under identical input conditions, since only one such amplifier was available. 4.7 Data Acquisition The strain and velocity signals were acquired and digitized using a high-speed, USB-based data acquisition board (USB 1604HS, Measurement Computing, Norton, MA). The signals were simultaneously sampled at 250 kHz. Sampling was triggered at a specified voltage on the rising slope of the velocity signal, and the board's pre-trigger sampling capability was used to capture the signals over a short duration preceding the trigger event. The board was programmed using LabVIEW and the ULxTM library of virtual instruments (VI's) supplied with the board. 4.8 Experimental Procedures 4.8.1 Input Velocity The input to the model was a step displacement applied by the hydraulic actuator to the base plate and acetabular subassembly. However, since no real system can supply an instantaneous displacement, and since it was vital to know the input accurately for subsequent use in the finite element analysis, the actual input was carefully measured. First, the MTS controller was tuned. The system was set to run a displacement square wave of an appropriate magnitude (1 mm), and the displacement signal was monitored with an oscilloscope. The controller's PID settings were then adjusted to maximize the displacement rate without causing overshoot. In this and in the experiments, the MTS actuator's LVDT displacement sensor was scaled into a short range of ±10 mm to improve its accuracy in the short displacement scenario of these tests. 51 After controller tuning, the BMR model was assembled in entirety as if for an actual experiment trial. The experimental procedure (as described below) was performed, but instead of recording femoral head velocity, the velocity of the base plate was measured with the laser vibrometer. A reduction trial was performed with the same displacement used in the actual reduction experiment, and the base-plate velocity was recorded with a digital oscilloscope. The actual loading during heel strike as per ISO standard is downward [1]. As the experimental BMR model is in an anatomically inverted orientation (as described in Chapter CHAPTER 2), the velocity input in the BMR model used to simulate the loading during heel strike phase is vertically up. This dynamic velocity record was applied subsequently as the input to the finite element simulation of the BMR model, as described in Chapter 5. 4.8.2 Reduction Trials The experiments were designed to have initial polar separation gap (PSG, Fig 7) values of 2, 3, and 4 mm in separate trials. Since these PSG dimensions could not be measured directly, the set-up conditions to establish them were determined by a partly trial-and-error, partly deductive process. First, the entire BMR model was set up such that the femoral head was fully reduced in the acetabular liner. This was after lubricating the bearings with a few drops of a bovine serum solution (serum with 30 mg/L protein content from Thermo Fisher Scientific, Logan, UT, diluted 50% with de-ionized water). The horizontal distance (x) of the femoral head relative to the edge of the acetabular block was measured using gauge blocks (arrangement shown in Fig 27). Then, the actuator (and entire pelvic subassembly) was lowered by an initial trial distance (y) such as 0.8 mm, whereupon the femur could rotate slightly to create edge contact (between 52 head and liner), accompanied by a reduction in x. The trial distance y is termed as drop distance because the actuator was lowered (dropped) by that distance. Then, applying the 105 N lateral spring force reduced x by a further small amount (e.g. 0.1 mm) due to compliance in the assembly. An accurately dimensioned 2D CAD model of the femoral construct, head, and liner was previously constructed, and from this, it was estimated that certain values of x would correspond with the desired PSG values. Therefore, the actuator was then lowered further from the trial distance y until the measured distance x (again measured with gauge blocks) diminished to the value obtained from the CAD model. The result was called the "actuator drop distance", and this set-up parameter was used to position the femoral head relative to the acetabular liner in the reduction trials. This procedure was repeated for each PSG value. To set up a reduction trial, the femoral and pelvic subassemblies were lubricated with the serum solution and positioned with the head fully reduced. Then, the actuator was lowered by the required drop distance. One of the strain gages was connected with the amplifier, and the bridge output was zeroed. Then, the 105N lateral spring force was applied. The laser head of the vibrometer was moved into position, leveled, and focused onto the retro-reflective tape on the femoral head. Then, the trial was performed by delivering a step displacement with the MTS actuator. The magnitude of the step displacement was slightly less (by 0.05 mm) than the drop distance; this was a precaution to avoid a potentially destructive crash between the femoral head and acetabular liner. At least three trials were performed for each PSG value. Trials were then repeated to acquire measurements from the other strain gage. 53 load cell adapter plate u-bracket rotation axis spring mechanism laser vibrometer pelvic subassembly actuator base plate sine plate Fig 21: Experimental setup of the BMR model 54 Fig 22: Proximal end of the femur construct; assembly of synthetic femur and implants - femoral stem and femoral head; the femur's greater trochanter was resected to provide a line of sight for the laser velocity measurement described in Section 4.2 Table 7: Material properties of the components of the BMR model Component Material Density (kg/m3) Elastic Modulus (GPa) Poisson's ratio Femur1 Cortical shell Short fiber filled epoxy 1640 13.7 0.42 Cancellous bone Polyurethane foam 270 0.155 0.35 Implants Femoral Head2 Al2O3 3960 380 0.26 Acetabular Liner2 Al2O3 3960 380 0.26 Acetabular Shell3 Ti-6Al-4V 4430 113.8 0.342 Femoral Stem3 Ti-6Al-4V 4430 113.8 0.342 Model pelvis1 Test block Polyurethane foam 480 0.462 0.3 1 Pacific Research Labs, Inc., Vashon, WA 2 [33] 3 www.matweb.com as viewed on Oct 6, 2010 55 Fig 23: Distal end of the femur construct with a press-fit dowel pin between the condyles 56 Spring system Hook Adjustment nut Connecting rod Rotation block Fig 24: Compression spring system (top) connected to the femur by means of a hook (bottom) 57 . Spring Load cell Rotation block Laser focus point a) b) Fig 25: Components of the spring system (spring in green, load cell and rotation block) and laser focus point (red spot) from the vibrometer focused on the femoral head to measure the velocity Fig 26: Strain gages installed on a) the femur and b) the femoral neck 58 Lateral side of the femoral head Gage block stack Reference edge of the test block Fig 27: Arrangement of gage blocks to measure the horizontal displacement of the femoral head CHAPTER 5 FINITE ELEMENT ANALYSIS OF BMR MODEL3 5.1 Introduction This chapter describes the finite-element analysis (FEA) of the BMR model and an aggressive submodeling technique used to quantify the contact force and stresses in the COC hip implants during sudden reduction. Using the Abaqus finite-element code [50], the solution strategy was divided into two key steps. First, the entire model was analyzed under inputs matching those recorded in one reduction trial; this step was called the global model analysis. Second, the model was partitioned to remove the majority of its geometry, other than small portions of the head and liner immediately surrounding the edge-loading contact point, and then this smaller model was analyzed with a refined mesh; this step was called the submodel analysis. The experimentally measured response parameters (femoral head velocity and femur strains) were used to validate the global model. The submodel analysis provided key required outputs, namely contact force and contact stresses that could not be measured directly in the experiment. Although the experimental BMR model was tested in an inverted anatomic orientation (with the pelvic subassembly beneath the femoral subassembly), the FEA was performed on a BMR model oriented in an anatomically upright orientation. This 3 The Abaqus analyses described in this chapter were performed by contractor Exponent, Inc., under our supervision. approach facilitated model and results interpretation and communication. The orientation and anatomic descriptions in this chapter are in accordance with the upright orientation. 5.2 Three-Dimensional Solid Model and Mesh Generation Three-dimensional (3D) CAD solid models of each component of the BMR assembly provided the geometry needed for the FEA. A CAD model of the 4GCF was acquired from its manufacturer. According to the manufacturer, the CAD model was based on data obtained from computerized tomography (CT) scans of a sample 4GCF. The femoral stem's CAD model was obtained from the manufacturer. The femoral head CAD model was built in SolidWorks using dimensions from its manufacturing drawings. The acetabular liner CAD model was built in Pro/Engineer (Pro/E) using dimensions from two sources. The contour of the spherical bearing surface (the ID) and the adjacent edge were digitized in 0.1 mm increments using a scanning touch probe CMM (Contoura, Zeiss, Germany). The 2D coordinate data points were merged in Pro/E with 2D sketch entities representing the liner's face and all its exterior features, and the resulting sketch was revolved to form an axisymmetric 3D model. We used measured ID and edge dimensions rather than nominal manufacturing drawing dimensions, because prior observations showed that actual dimensions, particularly near to the edge, could vary substantially from the nominal ones. Moreover, the edge contour would expectedly play a significant role in the edge-load forces and stresses during sudden reduction; so, accurate CAD model dimensions were vital. The acetabular shell was a simple Ø62 mm hemispherical shell with an internal Morse taper exactly matching that of the liner; its CAD model was created in Pro/E. The test block's CAD model was also built in Pro/E, and its Ø62 mm hemispherical bore was sized to exactly match the shell's outer diameter. 61 To create the virtual BMR model assembly, the CAD models were imported into Hypermesh (Altair Engineering, Troy, MI) and assembled as dictated by their mating surfaces, except that positioning the stem relative to the femur required extra steps. An initial measurement of the experimental femur construct provided data for a trial position in the Hypermesh assembly. This measurement was performed using a CMM, and it provided two data sets needed for the virtual assembly: (1) the position and the orientation of the distal pin relative to the femur and the CMM's reference frame, and (2) the position of the head relative to the pin and the reference frame. A detailed description of the CMM measurement scheme is given in Appendix B. The CMM measurement results were replicated in the virtual Hypermesh assembly of femur and stem. However, the initial virtual assembly showed that the stem interfered slightly with the femur, so it was judged that the CMM measurements did not position the stem relative to the femur accurately enough. Therefore, A/P and M/L radiographs of the femur construct were obtained. These were then used as visual guides while the stem was adjusted in the assembly to match its appearance (i.e., its position relative to the femur) in the radiographs. Where the stem geometry still interfered with the femur geometry, such as in the femur's intramedullary cancellous bone volume, the femur geometry was eliminated by Boolean subtraction. The femoral and pelvic subassemblies were then positioned with respect to each other as described in Chapter 2. The meshes generated for all the components of the virtual BMR model are shown in Fig 28-31. Each component was individually meshed using Hypermesh. Due to their complicated shapes, the stem and femur were meshed using 10-node linear tetrahedral elements (called C3D10M in Abaqus) generated by Hypermesh's automatic mesh 62 generator. To represent a press-fit, the interface between the stem and femur meshes was constrained using surface-to-surface constraints, which imposed a restriction that the two components could not slide relative to one another. The surfaces of the femur and the femoral stem were more densely meshed than the inner cores of these components as shown on the right side of Fig 30. This was done to reduce the computation time, and it was judged that this strategy would not affect the principal response parameters such as the strains on the femur construct and the stresses at the edge loading contact. The femoral head and the acetabular liner were meshed mainly with 8-node linear hexahedral elements (called C3D8R in Abaqus) in a rotational sweep. Both of these components had a locally refined mesh with an element edge length of approximately 0.1 mm near the edge load contact point (as shown in Fig 31) and, the largest element size away from the contact point was 2.6 mm for the femoral head and 3.6 mm for the liner. The acetabular shell and the test block were meshed with mainly tetrahedral elements. As with the femur-stem assembly, all of the other press-fit mating interfaces were modeled with surface-to-surface constraints. The head-liner contact interface, on the other hand, allowed friction, as described below. The meshes were exported from Hypermesh and imported into Abaqus for the computational analysis. 5.3 Material Properties All of the material properties were modeled as linear elastic, with values per Table 7, except for those of the femoral cortex. The femoral cortex was also modeled as a linear elastic material, but the appearance of error in initial model results (judged by comparison with experimental results), and a review of composite femur property data in relevant scientific literature [51], suggested that the manufacturer-provided properties 63 could be inaccurate. Therefore a separate experimental trial was performed to calibrate the femoral cortex properties. A second strain gage was installed on the femur's medial diaphysis, immediately below the first gage and oriented perpendicular to it (Fig 33). The femur construct was suspended vertically by the U-bracket from the load cell of a separate test frame. The test frame and femur were set up to compress the femur axially, with the femoral head contacting a hardened steel compression platen. The femoral head was positioned underneath and concentric to the load cell. The head was constrained from sliding by steel blocks clamped to the compression platen, and the platen near the head was greased to minimize friction. Then, the test frame crosshead was lowered at a constant rate of 0.36 in./min. to a total displacement of 0.03 in. while the strain in one gage was recorded simultaneously with the load. A second, identically configured trial was performed to record the strain in the transverse gage, though only after an approximately 45 minute delay, since the femur exhibited viscoelastic relaxation characteristics, as evidenced by a residual, slowly decreasing strain observed in the initially measured gage. The load measured in this trial was applied as input in a quasi-static finite element analysis of the femoral construct constrained and loaded to mimic the trial. When the model was analyzed using the manufacturer-supplied femoral cortex properties, the strains from the FEA did not match those from the trial, as shown in Fig 34. So, the model's material properties were scaled and the analysis was re-run. The results from this second analysis showed close agreement between measured and predicted strains (Fig 34), which validated the adjusted values of the femur's elastic properties. The elastic 64 modulus of 13.25 GPa and the Poisson's ratio of 0.41 determined in this calibration process were used in all subsequent BMR model analyses. The use of a linear-elastic material model for the femoral cortex, despite the viscoelastic behavior observed in this slow-loading trial, was a simplifying assumption. The simplification was initially rationalized on the basis of its common use in prior FEA studies involving composite femurs [51]. As described in further detail in the Discussion chapter, the retrospective justification for using assuming elasticity in the fast-loading study is simply that dynamic loading corresponds to a large Deborah number, for which material response is essentially elastic. 5.4 Loading and Boundary Conditions The boundary conditions (BCs) employed in the FEA of the BMR model represented the BCs described in Chapter 2. The femur's distal pin axis was represented by a cylindrical hole passing through the distal femur in the same location as the physical distal pin. The nodes on the surface of this hole were constrained to only rotate about an axis coaxial to the hole. Otherwise, the pin itself was not modeled as a distinct 3D entity in the FE model. Realistic compliance at the pin axis was modeled by implementing horizontal and vertical linear spring elements with spring constants matching the effective fixture stiffness values measured as described in Section 4.3. The proximal surface of the pelvic test block was constrained to translate only in the vertical direction. A tensile spring element with a spring constant of 8.88 N/mm was incorporated into the FEA to represent the experiment's lateral spring. Though the experiment used a compression spring, the spring mechanism transformed its compression into a force that pulled laterally on the femur, thus the use of a tensile spring element in the FEA. The 65 spring constant was determined by direct force-displacement calibration of the experimental spring, using a separate test machine. The loading and BCs applied in the FE version of the BMR model are illustrated in Fig 35. To commence the global model analysis, the pelvic subassembly was raised vertically to mimic the drop distance applied in the experiment. Then, the load inputs were applied in two steps. The first was a static equilibrium step, in which a static pre-load was applied via the lateral spring element; the solution in this step was obtained using Abaqus' implicit solver (Abaqus/Standard). This step produced a small femoral head displacement that was compared against its corresponding value measured in the experiment. The end of the static step represented a condition of edge loading with the femoral head subluxed, at a time just prior to heel strike and subsequent head reduction. The second load step comprised the dynamic analysis of sudden reduction; the solution in this step was obtained using Abaqus' explicit solver (Abaqus/Explicit). The dynamic base plate velocity recorded in the reduction experiment was applied as an input across the test block's proximal surface. This input forced the femur to rotate and the head to reduce into the liner, thus creating a simulation of the experiment. 5.5 Contact Modeling For the static step using Abaqus/Standard, a surface-to-surface contact algorithm was used. A pure master-slave relationship was used, with the femoral head as the master and the acetabular liner as the slave. To enforce contact between the femoral head and the acetabular liner, hard pressure-overclosure behavior with a linear penalty method was chosen. Frictional constraints (μ=0.06) were enforced with a penalty method. For the 66 dynamic step using Abaqus/Explicit, a node-to-surface contact algorithm was used with a balanced master-slave relationship, which completely prohibits material interpenetration. 5.6 Submodeling of BMR Model Attempting to run a fully or a locally refined global model with explicit time integration is a high runtime consumption process as it involves analyzing a full scale model. To overcome this, an aggressive submodeling approach was performed allowing for fine mesh densities and a greatly reduced model volume, which gives accuracy and computational efficiency. The submodeling technique undertaken in this project is aggressive because the volume of the submodel is less than 1/100th of that of the global model. The submodel comprises portions of the components immediately surrounding the contact region as described later in this section. Elkins et al. [52] used submodeling in the field of orthopedics to analyze the stress concentrations on COC hip bearing couple. That submodel was an annulus (portion of the acetabular liner) with an element size of 0.3 mm whereas the BMR global model was locally refined with an element size of about 0.1 mm. Cormier et al. analyzed a 2D static contact problem of turbine blades using an aggressive submodeling approach [42]. Submodeling analysis of the BMR model extends the submodeling application to a 3D dynamic contact problem. Elkins et al. [52] employed a conservative submodeling approach in the sense that their submodel size was large in comparison to the contact area; an aggressive, carefully executed, submodeling approach using a smaller submodel volume could have reduced computation costs without adversely affecting accuracy. The BMR model analysis included a submodeling technique in which the submodel is much 67 smaller than the submodel in [52]. This section is dedicated to describing this aggressive submodeling approach undertaken to analyze the BMR model. 5.6.1 Submodel Generation The BMR submodel, as shown in Fig 36b, was a small region surrounding the contact area (where the edge loading occurred). The submodel was partitioned from the global model components that were members in the contact, namely the femoral head and the acetabular liner. The geometries of the two components were partitioned as a sector with included angles of 80° (labeled in Fig 36c) in an inclined transverse plane and 20° in the frontal plane (labeled in Section A-A of Fig 36c). Then, a thickness of 3.5 mm in each of the components was retained, which is the submodel, and the rest of the geometry was removed. 5.6.2 Boundary Conditions The BCs on this submodel were enforced by interpolating the displacements taken from the global model at those interfaces where the submodel was partitioned (the "Cuts" in Fig 36c) from the global model. The interpolation was done within Abaqus. 5.6.3 Submodeling Analysis Procedure Submodeling is based on Saint Venant's principle [53], which allows modeling only a portion of an entire structure by applying equivalent loads to the boundary of a subbody and as long as the BCs on the subbody roughly approximate the actual tractions on that boundary, the details of the stress fields away from the boundary are modeled accurately. The creation of the BMR submodel geometry (described in Section 5.6.1) is governed by this principle, because an equivalent load system is applied to the submodel 68 geometry as existed in the global model, and the boundary of the submodel is relatively far away from the region of contact. So, the submodel partitions from the global model were made around the contact area at a distance where the stress gradients were negligible. The submodel was meshed with 10-node hexahedral elements, and the mesh was much finer than the global mesh both at the contact and in the region surrounding the contact. The element edge length at the contact was 0.01 mm (1/10th of the global mesh element edge length at the contact) for both the femoral head and acetabular liner submodels. The largest element edge length was 1.17 mm in the head portion and 1.15 mm in the liner portion. Then, the displacement BCs were imposed on the submodel and the BMR submodel was solved in Abaqus. Only a single locally refined submodel was analyzed. This procedure was adapted to analyze the BMR model because (1) it was expected that the contact stresses predicted by the global model analysis would be poorly resolved, since the global model's elements near the contact region were close in size (e.g., 0.1 mm) to the minor dimension (e.g., <0.3 mm) of experimentally observed edge-loading contact patches in COC hip bearings [54] and, (2) the computation time to solve the global model itself was ~31 hours with a 12-cpu machine (Intel Xeon X5670, 2.93 GHz) with 48 GB memory. 69 Fig 28: Mesh of the femur (left) and zoomed-in view (right) showing the tetrahedral mesh density Fig 29: Mesh of the femoral stem (left) and zoomed-in view (right) showing the tetrahedral mesh density 70 Interfered femur geometry eliminated Femoral Stem Cancellous bone Cortical bone c) b) a) d) Fig 31: Locally refined hexahedral meshes of a) femoral head, b) acetabular liner; c) Zoomed in views of the local refinement; d) Tetrahedral mesh of the acetabular shell and the test block Fig 30: Femoral stem and cancellous bone assembly (left) showing the penetration of the stem mesh into the cancellous mesh; (right) Section view of assembly of femoral stem, cancellous and cortical bones showing the inner core mesh density 71 Longitudinal strain gage Lateral strain gage PSG Fig 32: Section view of femoral head and acetabular liner meshes and the PSG (left); Detail view showing a locally refined mesh at the contact (right) Fig 33: Two strain gages installed on the femur's medial diaphysis. One strain gage measured longitudinal strain and the other measured lateral strain 72 Fig 34: Longitudinal (Parallel) strain and Lateral (Transverse) strain vs. force using the original and the adjusted material properties; the strains obtained using adjusted material properties match with the experimentally obtained strains. 73 Velocity input Lateral spring force Rotation axis Y X Fig 35: Illustration of the loading and the BCs in the FE version of the BMR model (shown in section view) 74 Head Submodel a) Liner Submodel BMR Submodel b) c) Fig 36: Meshes of the submodel components, refined at the contact region (top left); The BMR submodel (top right); Illustration of the submodel generation along with its dimensional details. "Cut" labels indicate where the submodel's geometry was partitioned from the global model's geometry (Dimensions are in mm.) CHAPTER 6 RESULTS: TESTING AND ANALYSIS OF BMR MODEL The results in this chapter illustrate the rapid structural responses in the femur and the femoral neck and, the contact mechanics of the bearing couples during the sudden femoral head reduction event as analyzed experimentally and computationally. 6.1 Experimental Results The experiment was conducted for three PSG values: 2 mm, 3 mm and 4 mm. The following subsections give the experimental comparison of different responses of the BMR model for the three PSGs. It was expected that the magnitude of the responses of the model would increase as the PSG increased. 6.1.1 Velocity of the Femoral Head The velocity of the femoral head for the 2 mm PSG as measured by a 3D laser vibrometer is given in Fig 37a. The velocity of the femoral head measured for the three PSG values is shown in Fig 37b, which shows that an increase in PSG results in higher magnitude of the peak velocity of (and hence, a higher contact force on) the femoral head. Fig 37b shows a 12 ms time span of data; it encompasses the period of interest, when the head and the liner undergo edge loading. 76 6.1.2 Strain in the Femur and the Femoral Neck Fig 38 (a and b) shows the measured strain on the compressive side of the components - the medial side of the femur and the medial side of the femoral neck respectively. Fig 38 (c and d) shows a comparison of the strains measured on the femur and the femoral neck respectively, for the three PSGs. The similarity in the initial strain values, for all the PSGs recorded for each of the femur and the femoral neck, results from the use of a consistent spring force value applied in the test's initial static load phase. As expected, the peak strain became higher as the PSG increased, following the same pattern as for the velocity of the femoral head. 6.2 Validation of the BMR Model The FE version of the BMR model was validated against the experimental version using three response parameters for the 2 mm PSG. The response parameters were (1) velocity of the femoral head during reduction, (2) strain on the femur and, (3) strain on the femoral neck. This section further discusses the validation of the BMR model. 6.2.1 Velocity of the Femoral Head A time history of velocity was extracted from the FEA simulation at a surface node located on the femoral head in approximately the same location that the laser sensed in the experiment. As shown in Fig 25, the velocity of the femoral head in the experiment was recorded by focusing the laser on approximately the farthest point located on the lateral side of the femoral head after applying the spring force. The FEA data were extracted from the surface node that was on the farthest lateral side of the femoral head at the end of the static load step. 77 Fig 39a compares the velocity of the femoral head in the FEA and the experiment. The peak velocity of the femoral head in the FEA showed a 25% error when compared against the experimental velocity. Also, the femoral head seemed to be moving at a faster rate during the reduction event when compared to the experiment, resulting in a higher peak velocity. The differences in the peak velocities and the rates at which the femoral head reduced can be attributed to a difference in the friction value at the contact. This is further discussed in the Discussion chapter. Fig 39b shows a focused view of the velocity of the femoral head for a time span of the initial 12 ms. It shows a better view of the velocity for the period of interest, i.e., the initial 8 ms during which the femoral head reduction happened. After this time interval, rapid oscillations in the FE femoral head velocity curve are seen in contrast to the gradual experimental curve. These oscillations can be attributed to the femoral head's bouncing around inside of the liner ID. Also, the FEA did not model the presence of a small amount of lubricant within the liner, which was present in the experiment; that lubricant in the experiment (not modeled in the FEA) might have prevented or damped out such rapid oscillations. In the FEA, it appears that the head collided with the liner ID first at 8 ms into the simulation. In the experiment, no such definitive collision can be discerned, probably because the head actually collided with the fluid within the liner that became pressurized very rapidly, since the clearance of the bearings is very small (<60 μm). This explanation is supported by the observation that fluid splashed from the liner onto the components in some experimental trials. Nevertheless, these oscillations are of little concern, since it happens after the period of interest (period of edge-loading) has ended. 78 6.2.2 Strain in the Femur and the Femoral Neck Fig 40 (a and b) shows the comparison of the FE and the experimental strains on the femur and the femoral neck, respectively, both for the total time span for which the experimental data were recorded. Initially, the FEA strains show good agreement with the experimental strains though, later during the event, the magnitudes of the FE strain peaks and rates are lower than the magnitudes of experimental peaks and rates, with a 7% error in the strain on the femur and an 8% error in the strain on the femoral neck. This shows that the femur in the FEA did not bend as far as the femur in the experiment. Fig 40 (c and d) shows a focused view of the strains of the femur and the femoral neck, respectively, after the initial 8 ms. The curve after ~8 ms has local peaks. These peaks are seen at intervals of the transit time of the stress waves initiated due to an impact of the femoral head on the inside (towards the pole) of the acetabular liner. This interesting phenomenon is further described in the Discussion chapter. As seen in the velocity plots, similar rapid oscillations are seen for the strain on the femoral neck after the period of interest. Since the period of interest ended before the rapid oscillations occurred, they can be ignored. 6.3 Inferred Outputs from Global Model and Submodel Analyses This section provides the inferred outputs from the FEA of the BMR model (inferred outputs are those that do not have corresponding experimental measurements). FEA was performed for a single PSG (2 mm). Contact force (Fig 41), contact stresses (Fig 42) and, contact areas (Fig 43) for each of the head and the liner are the inferred outputs from the FEA for the 2 mm PSG. Global model and submodel results of these outputs have also been compared against each other and are discussed in brief. 79 The contact force and the contact area are defined in Abaqus as follows: Contact Force: The contact force is the total normal force due to contact pressure. The Abaqus keyword used to extract the contact force is CFN. Contact Area: The total contact area is defined as the sum of area of all the facets where there is contact force. The Abaqus keyword used here to extract the contact area is CAREA. The global model and submodel contact force and contact stress plots show a large magnitude of error when compared to each other. Given 25% error in the measured quantities (i.e., velocity and strain as shown in Section 6.2) used to validate the FE model, the global model contact force clearly has considerable uncertainty. Further investigation on the source of error for the velocity of the femoral head has to be done to see the effect of change in velocity on the contact force as described in the Discussion chapter. This error is attributed to using displacement BCs and to simultaneously changing the contact stiffness by altering the mesh density in the submodel. The results might be improved with traction BCs. This source of error is discussed in detail in the Discussion chapter. 6.4 Contour Plots This section shows contour plots of stress at the contact on both the head and the liner in the global and the submodel analyses and the strain contour plots on the femur and the femoral neck. The contour plots shown in this section correspond to a time point of 6.9 ms into the event. Fig 44 shows the stress distribution at the contact on the acetabular liner and the femoral head respectively, from the global model analysis. These results show the 80 location of the contact area on the rim of the acetabular liner and the femoral head qualitatively. At about 7 ms into the simulation (which is clearly evident in animations of the FE simulation showing the contact stresses), the contact position changes rapidly from being on the crest to a position away from the rim of the liner suggesting the contact stresses increases causing severe wear. Fig 45 shows the stress distribution on the liner and the head respectively, from the submodel analysis. Fig 46 shows the stress distribution in the femur (section view) at 20 X magnitude. The stress on the compressive side of the femur-femoral stem interface is maximum compared to the stresses in the femur, suggesting that the design of the femoral stem may be revised further to reduce these stresses during the reduction. Fig 47 shows the contour plots of the minimum principal strain on the medial side of the femur and the femoral neck. 81 Fig 37: Velocity response of the femoral head during the 2 mm PSG reduction event (top);Overlay of the velocity of the femoral head for the three PSGs shown for a time span of 12 ms to better show the responses during the reduction event (bottom) 82 a) b) Fig 38: Measured strain response for 2 mm PSG in a) the femur, b) the femoral neck over a total span of 0.1s; Overlay of the strain response for the three PSGs in c) the femur and d) the femoral neck over a span of 12ms 83 a) b) Fig 39: Comparison of velocity of the femoral head during the reduction event for 2 mm PSG using FEA and experiment a) for the total time span for which experimental data was collected and, b) focused view for 12 ms 84 a) b) c) d) Fig 40: Comparison of strain for 2 mm PSG from FEA and experiment in a) the femur, b) the femoral neck, both for the total time span for which the experimental data was recorded; focused views of the strain comparison for 12 ms for c) the femur and, d) the femoral neck 85 Fig 41: Comparison of the global model and the submodel contact force 0 1 2 3 4 5 6 7 8 0 200 400 600 800 1000 1200 1400 Time (ms) Contact Force (N) Global Model Submodel 86 Fig 42: Comparison of the global model and the submodel results of contact pressure for a) the femoral head and, b) the acetabular liner 0 1 2 3 4 5 6 7 8 0 100 200 300 400 500 600 700 800 Time (ms) Contact Pressure (MPa) Global Model Submodel 0 1 2 3 4 5 6 7 8 0 200 400 600 800 1000 1200 Time (ms) Contact Pressure (MPa) Global Model Submodel a) b) 87 Fig 43: Comparison of the global model and the submodel results of contact area for a) the femoral head and, b) the acetabular liner 0 1 2 3 4 5 6 7 8 0 1 2 3 4 5 6 Time (ms) Contact Area (mm2) Global Model Submodel 0 1 2 3 4 5 6 7 8 0 0.5 1 1.5 2 2.5 3 3.5 4 Time (ms) Contact Area (mm2) Global Model Submodel a) b) 88 Fig 44: Contact stress distribution during the reduction event at 6.9 ms for the liner (top) and, the head (bottom) from the global model analysis 89 Fig 45: Contact stress distribution during the reduction event at 6.9 ms for the liner (top) and, the head (bottom) from the submodel analysis 90 Fig 46: Stress distribution in the femoral subassembly (section view) at the time of maximum stress at the contact; displacements shown at 20 X magnitude Fig 47: Contour plots of minimum principal strain on a point (where a strain gage was installed to measure the longitudinal strain) on the femur (top) and on the femoral stem (bottom) CHAPTER 7 DISCUSSION 7.1 Validation of the PSG in the FE Model Validation of the PSG in the BMR model is important because the position of the initial point of head-liner contact is governed by this gap. The inner profile of the liner in the region of the rim has several differently shaped features - including a sharp edge called a crest, a straight segment about 1 mm long, and the edge radius. The PSG determines which of these features, the head initially contacts, and the slope at the point of contact strongly affects the moment arm length of the contact force. Since a key objective of the project was to accurately analyze the contact force, it was essential to set up the PSG accurately, both experimentally and computationally. In the BMR model, the femoral head had three positions: (1) an initial fully reduced position (clearance included), (2) an intermediate position that mimicked the drop distance in the experiment, and (3) a final subluxed position with the desired PSG. The total horizontal displacement covered by the farthest point on the medial side of the femoral head from position (1) to (3) was measured computationally and compared against that measured in the experiment (as described in Section 4.8.2). The horizontal distance between positions (1) and (2) was measured from the 2D model and the FEA was set up initially starting with the position (2). The horizontal displacement between positions (2) and (3) was measured from the FE model (which constitutes the static load 92 phase). These two distances added together gave the total horizontal displacement in the FE model. For the 2 mm PSG, the horizontal displacement was 1.56 mm measured experimentally and 1.37 mm measured computationally. Hence, the head was more subluxed in the experiment. Since the separate PSGs examined experimentally were only 1 mm different from each other, the end results from different PSGs may have been confused with one another as to which PSG the end results belong to. Hence, implementation of this deliberate setup procedure is important. Also, the verification of the PSG is an additional confirmation of the initial state of the femur at the end of the static load phase. As the femur exhibited viscoelastic relaxation characteristics (discussed in detail below), it is important to account for the initial deformation under static loads before applying the dynamic loads. This verification combined with the verification of the initial strain on the femur and the femoral neck confirms the position of the components of the model before applying the dynamic input. 7.2 Behavior of the Femoral Cortex Material Composite femurs, which behave close to human femurs, have been preferred for biomechanical testing by researchers. In FEA involving composite femurs, researchers have modeled both the cancellous and the cortical materials as linear elastic and isotropic [51]. In contrast, during the axial compressive load test (described in Section 5.3), the femoral cortex exhibited viscoelastic relaxation characteristics. Similar characteristics were observed during the BMR test. The strain recorded on the femur experienced damping over a period of time after the reduction event. As seen in Fig 38a, the magnitude of the peak strain decreased gradually after the period of interest. Nonetheless, 93 the simplification that the femur can be modeled as linear elastic and isotropic in the FEA of the BMR model is justified, because the time period of interest was 5 ms, during which the femur behaves approximately as a linear elastic and isotropic material. 7.3 Effect of Friction This section discusses the effect of friction on (1) the PSG, (2) the velocity of the femoral head, and (3) the contact force between the bearing surfaces. Section 7.3.1 describes the effect of different lubricants that changed the coefficient of friction (CoF) at the contact. Sections 7.3.2 and 7.3.3 describe the effect of CoF on the velocity of the femoral head and the contact force respectively, as analyzed from the rigid body analysis. 7.3.1 Effect of Lubrication on the PSG Lubrication applied on the contacting surfaces was important in the BMR test because it contributed to the slipping of the femoral head on the rim of the acetabular liner. The slipping effect, affected by the CoF, changes the PSG. The lubrication fluid used in the BMR test was 50% diluted bovine serum, which corresponds to a CoF of 0.06 according to Scholes et al [45]. As reported in the same paper, the coefficient of friction between COC bearing surfaces increased as the % of dilution increased. A similar pattern was observed during the BMR test. The reduction test was performed using (1) 50% diluted and, (2) undiluted bovine serum. The horizontal displacement (described in Section 7.1) measured in both the cases, was found to be less in case (2) than in case (1), showing that the femoral head was sliding out more when the lubricant was less dilute. The decrease in the slipping effect (affected by the CoF) as the % of dilution increased, can be attributed to the decrease in the protein content due to dilution [45]. 94 7.3.2 Effect of Friction on the Velocity of the Femoral Head The CoF played an important role in affecting both the static and the dynamic responses of the BMR model. Therefore, error in the assumed CoF value for the FEA was regarded as an important potential root cause of the differences between the FEA and the experimental results. This section describes additional analysis to show the effects of friction on the BMR model. The rigid body (described in Section 2.6) was analyzed to estimate the effect of the CoF on the velocity of the femoral head. Eq. 7, obtained by integrating the acceleration term in Eq. 5, gives the velocity of the femoral head during the reduction event. π£ = ττ«∫ πΌ π‘ 0 ππ‘ττ¨ (π + π) = (π−ππ‘)πΆπ− π πΆπ+ π₯π πΌπ (π + π) (7) The time record of the velocity (v) plotted for a range of CoFs is shown in Fig 48. As the CoF increased, the velocity of the femoral head decreased. It can also be observed that as the CoF increased, the acceleration of the femoral head decreased. According to this simple analysis, the CoF has a linear effect on the femoral head velocity. As seen in Fig 37, the peak velocity of the femoral head recorded in the experiment was less than that in the FEA. Therefore, we adjusted the CoF value in the FEA, using predictions from this simple analysis. So, the FEA of the BMR global model was rerun with a CoF value of 0.11 (approximately doubling the assumed CoF). Yet the results remained unch |