Title | Spontaneous Venous Pulsations Detected With Infrared Videography |
Creator | James A. McHugh; Linda DʼAntona; Ahmed K. Toma; Fion D. Bremner |
Affiliation | Department of Neuro-Ophthalmology (JAM, FDB), National Hospital for Neurology and Neurosurgery, London, United Kingdom; Department of Ophthalmology (JAM), King's College Hospital London, United Kingdom; and Department of Neurosurgery (LD, AKT), National Hospital for Neurology and Neurosurgery, London, United Kingdom |
Abstract | Background: Assessment of spontaneous venous pulsation (SVP) is commonly undertaken to help determine whether intracranial pressure (ICP) is elevated. Previous studies using direct ophthalmoscopy or slit-lamp assessments have found that SVP is not observed in 67%-81% of subjects with normal ICP, and that interobserver agreement when grading SVP is poor. Methods: Patients (n = 105) undergoing clinically indicated retinal OCT scans, who were all believed to have normal ICP, had 10-second infrared video recordings performed with the Heidelberg Spectralis OCT system (Heidelberg Engineering GmbH, Heidelberg, Germany). The presence and amplitude of SVP in each video was independently graded by 2 neuro-ophthalmologists. Results: The 2 observers found SVP present in 97% and 98% of right eyes and in one or both eyes in 99% and 100% of subjects. Interobserver agreement was high (Cohen's kappa 0.82 for right eyes). Optic discs with a smaller cup had a significantly lower SVP amplitude (Spearman's rho = 0.22, P = 0.02). Conclusions: Infrared video is widely available in eye clinics by the use of OCT imaging systems and is substantially more sensitive in detecting SVP than traditional assessments using ophthalmoscopy. SVP is absent in as few as 1% of people with presumed normal ICP. |
Subject | Female; Humans; Intracranial Pressure / physiology; Intraocular Pressure; Male; Middle Aged; Ophthalmoscopy / methods; Ophthalmoscopy / mortality; Optic Disk / diagnostic imaging; Retinal Vein / diagnostic imaging; Tomography, Optical Coherence / methods; Video Recording / methods |
OCR Text | Show Original Contribution Spontaneous Venous Pulsations Detected With Infrared Videography James A. McHugh, MEd, FRCOphth, Linda D'Antona, MBBS, MD, Ahmed K. Toma, MD, FRCS, Fion D. Bremner Background: Assessment of spontaneous venous pulsation (SVP) is commonly undertaken to help determine whether intracranial pressure (ICP) is elevated. Previous studies using direct ophthalmoscopy or slit-lamp assessments have found that SVP is not observed in 67%-81% of subjects with normal ICP, and that interobserver agreement when grading SVP is poor. Methods: Patients (n = 105) undergoing clinically indicated retinal OCT scans, who were all believed to have normal ICP, had 10-second infrared video recordings performed with the Heidelberg Spectralis OCT system (Heidelberg Engineering GmbH, Heidelberg, Germany). The presence and amplitude of SVP in each video was independently graded by 2 neuroophthalmologists. Results: The 2 observers found SVP present in 97% and 98% of right eyes and in one or both eyes in 99% and 100% of subjects. Interobserver agreement was high (Cohen's kappa 0.82 for right eyes). Optic discs with a smaller cup had a significantly lower SVP amplitude (Spearman's rho = 0.22, P = 0.02). Conclusions: Infrared video is widely available in eye clinics by the use of OCT imaging systems and is substantially more sensitive in detecting SVP than traditional assessments using ophthalmoscopy. SVP is absent in as few as 1% of people with presumed normal ICP. Journal of Neuro-Ophthalmology 2020;40:174-177 doi: 10.1097/WNO.0000000000000815 © 2019 by North American Neuro-Ophthalmology Society E levated intracranial pressure (ICP) is a feature of various neurological conditions including idiopathic intracranial hypertension, hydrocephalus, intracranial mass lesions, Department of Neuro-Ophthalmology (JAM, FDB), National Hospital for Neurology and Neurosurgery, London, United Kingdom; Department of Ophthalmology (JAM), King's College Hospital London, United Kingdom; and Department of Neurosurgery (LD, AKT), National Hospital for Neurology and Neurosurgery, London, United Kingdom. The authors report no conflicts of interest. Supplemental digital content is available for this article. Direct URL citations appear in the printed text and are provided in the HTML and PDF versions of this article on the journal's Web site (www.jneuro-ophthalmology.com). Address correspondence to James A. McHugh, FRCOphth, Department of Ophthalmology, King's College Hospital, Denmark Hill, London SE5 9RS, London, United Kingdom; E-mail: James.mchugh1@nhs.net 174 trauma, stroke, and prolonged reduction of gravity in astronauts (1). The gold standard for ICP assessment is lumbar puncture opening pressure; however, many clinical centers now use continuous ICP monitoring with intraparenchymal pressure transducers. Both techniques are invasive and risk bleeding and infection (2,3). Owing to the risks of invasive ICP measurement, many attempts have been made to develop reliable noninvasive tests for raised ICP. Computer tomography and MRI can be used to study the diameter of the optic nerve sheath and other anatomical indicators of high ICP (4,5). Ultrasound can be used to measure the retrobulbar optic nerve sheath diameter (6), and transcranial doppler can measure blood flow in the basal cerebral arteries (7). Cochlear fluid pressure can be used as a surrogate measure of cerebrospinal fluid (CSF) pressure and can be estimated by tympanic membrane displacement or re-emitted sound in response to acoustic stimuli (8,9). Pupillometry gives prognostic information in traumatic brain injury, possibly mediated through elevated ICP (10). Abnormalities of electronically recorded eye tracking are associated with high ICP (11). Ophthalmoscopy is widely used to assess for papilledema, which usually develops within hours or days of developing high ICP. However, patients with pre-existing optic atrophy, and some with apparently healthy optic nerves, may not develop papilledema even with very high ICP (12). Each of these noninvasive techniques of ICP assessment has its limitations, variously including suboptimal sensitivity, inter-rater variability, significant variation between healthy individuals, expense, time required, radiation exposure, or a need for baseline invasive ICP measurements. For these reasons, none has been universally accepted as a screening tool, and lumbar puncture or invasive ICP measurements remain essential in patients in whom intracranial hypertension is suspected. Spontaneous retinal venous pulsation (SVP) assessment offers an alternative noninvasive approach to assessing ICP. By the early 20th century, evidence from animal studies had McHugh et al: J Neuro-Ophthalmol 2020; 40: 174-177 Copyright © North American Neuro-Ophthalmology Society. Unauthorized reproduction of this article is prohibited. Original Contribution clearly established that SVP disappeared when ICP was high and could be re-established by elevation of the intraocular pressure until it exceeded ICP by around 5-10 mm Hg (13). The mechanism of SVP has been studied in detail and a variety of mathematical models proposed (14,15). Venous pulsation synchronized with the cardiac cycle occurs in highly compliant (expansible) veins near the site of high venous pressure gradients. The main site at which this occurs in humans is the optic disc. On entering the disc, the central retinal vein's pressure is determined by intraocular pressure; as the vein passes through the sclera, its pressure decreases reflecting the lower pressure of the CSF around the retrobulbar optic nerve. Cessation of SVP in the presence of high ICP reflects the diminution or reversal of this physiological venous pressure gradient. The resulting engorgement of the retinal veins due to decreased outflow inhibits their ability to change diameter during the ocular pressure cycle (14,15). In humans, SVP is not observable in up to a third of healthy individuals when examined with either an ophthalmoscope or slit lamp (16,17,20). We have found that SVP is more easily visible using the en-face infrared image of the fundus seen during image acquisition with the Spectralis ocular coherence tomography (OCT) system (Heidelberg Engineering GmbH, Heidelberg, Germany). Some Spectralis models allow for short fundus infrared video segments of up to 30 seconds to be recorded using scanning laser ophthalmoscopy. Image stabilization can be applied to video segments using the Spectralis's inbuilt software before video playback, eliminating motion artefacts that could be mistaken for SVP, and making small-amplitude SVP readily visible. Use of the Spectralis system to identify SVP has previously been described by Kim et al, who assessed SVP in glaucoma subjects, with SVP graded in a binary fashion as present or absent. SVP was found to occur less commonly in patients with glaucomatous disc cupping (18,19). There are no previous descriptions of this assessment method in the neuro-ophthalmic literature. This article describes a study designed to establish the sensitivity of motion-stabilized infrared fundus video in detecting and grading SVP in patients without ICPaffecting disease and to establish the inter-rater reliability of a 4-point SVP grading scale using this technique. neuritis (without disc swelling), and hereditary ataxias. The mean age of participants was 47.8 ± 15.9 years. Most of the participants (69.5%) were women. All participants gave consent to participation in the study, which was approved by a research ethics committee. The Spectralis OCT system was used to record a 10-second infrared video of each fundus, centred on the optic disc (See Supplemental Digital Content, Video, http://links.lww.com/ WNO/A378). Each recording was viewed by 2 independent assessors (J.M. and F.B.). All videos were motion-stabilized using the Spectralis' inbuilt software. The presence and amplitude of venous pulsation were graded using the system described by Hedges et al and later used by Wong and White (17,20): Grade 0 = no SVP, Grade 1 = ,33% change in venous diameter, Grade 2 = between 33% and 67% change in venous diameter, and Grade 3 = .67% change in venous diameter. Each eye also had a structural OCT of the disc, including a measurement of the vertical cup:disc ratio at the level of the Bruch membrane. RESULTS Gradable infrared video recordings of the ocular fundus were possible in all 210 eyes examined. Observer 1 found METHODS Participants were recruited from patients undergoing routine OCT scanning in the eye clinic of the National Hospital for Neurology and Neurosurgery. Potential participants were excluded if they had optic disc swelling or pseudopapilledema, symptoms suggestive of high ICP, pathological disc cupping due to glaucoma, or a medical history of abnormal ICP. The most common indication for OCT was routine screening for hydroxychloroquine retinopathy (31.4%). Other participants had a variety of neuro-ophthalmic problems not associated with abnormal ICP, including optic neuropathies, optic McHugh et al: J Neuro-Ophthalmol 2020; 40: 174-177 FIG. 1. Frequency of different grades of spontaneous venous pulsation (SVP) in normal subjects. Black columns show data from the present study using infra‐red videography. Grey columns show data in patients with normal ICP from Wong & White (17) using direct ophthalmoscopy. 175 Copyright © North American Neuro-Ophthalmology Society. Unauthorized reproduction of this article is prohibited. Original Contribution TABLE 1. Right eye SVP grading by 2 independent observers, using the grading system described by Hedges et al (20) Observer 1 Observer 2 Grade 0 Grade 1 Grade 2 Grade 3 Grade Grade Grade Grade 1 1 0 0 2 15 1 0 0 12 39 3 0 1 7 23 0 1 2 3 SVP in 103/105 right eyes (98%) and SVP in at least one eye of 104/105 subjects (99%). Observer 2 found SVP in 102/105 right eyes (97%) and SVP in at least one eye of all 105 subjects (100%). The frequency of observing different grades of SVP using video recordings is shown in Figure 1. Cohen's kappa for SVP grades of subjects' right eyes between the 2 independent assessors was 0.82. When comparing the highest grade of SVP seen in either eye, Cohen's kappa between the 2 assessors was 0.89, indicating high levels of inter-rater agreement. Table 1 shows the SVP gradings assigned by each assessor to right eyes, demonstrating the relatively low number of discrepancies between the 2 observers. Figure 2 summarizes the relationship between SVP and the cup:disc ratio (CDR). Discs with a larger cup had higher-amplitude SVP (Spearman's rho 0.22, P = 0.02). However, even in the group of eyes with no cup (CDR zero) at the level of the Bruch membrane, 85% of eyes had detectable SVP using infrared video assessment. CONCLUSIONS This study demonstrates that in a group of outpatients attending a neuro-ophthalmology clinic, it was possible in every case to record infrared fundus videos of sufficient quality to allow for grading of SVP. Comparison of our results with earlier studies shows that SVP is visible in the vast majority (99%) of eyes of patients with normal ICP using infrared video, compared with 67%-81% of eyes in patients with normal ICP using a direct ophthalmoscope (16,17,20). Hedges et al (20) reported that discs with large cups and those where veins could be viewed end-on had higher rates of visible SVP (69%). Crowded discs without cups and those in which disc veins were obscured by arterioles or glial tissue had much lower rates of visible SVP (6%). The inclusion of arteriovenous relationships in their disc grading system prevents the relationship between cup size and SVP from being understood in isolation. Our study used the SVP grading system described by Hedges et al to analyze SVP amplitude, but discs were analyzed according to the cup:disc ratio. Our study showed a clear relationship between a smaller cup:disc ratio and higher proportion of eyes with low-amplitude or absent SVP. There was a much higher rate of SVP in all disc categories in our videographic study than in any of the disc categories reported by Hedges et al using conventional ophthalmoscopy (20). It can be inferred from our results that SVP is in fact present in the great majority of healthy subjects, and the significant minority in which they could not previously be detected is largely due to the inherent limitations of examination with an ophthalmoscope or slit lamp rather than a true absence of retinal venous pulsations in those patients. We suggest that the difference is primarily due to the ability of the Spectralis system to motion-stabilize recorded images, revealing subtle pulsations, which would otherwise be extremely difficult to observe. A further advantage of the Spectralis OCT in particular is that it FIG. 2. Spontaneous venous pulsation (SVP) grades of right eyes with different cup:disc ratios. 176 McHugh et al: J Neuro-Ophthalmol 2020; 40: 174-177 Copyright © North American Neuro-Ophthalmology Society. Unauthorized reproduction of this article is prohibited. Original Contribution uses a near infrared light source which generates less reflections from vessel walls than visible light and is displayed as a black-and-white image, compared with the red-on-tan of vessels seen against the retina and disc with visible light. Fundus video recordings can be magnified and each recording played as a loop to allow for a considered decision on whether pulsations are present. These technical advantages are also reflected in the much higher rate of agreement between the 2 assessors in this study: Cohen's kappa 0.82 in right eyes and 0.89 when comparing the higher value of the 2 eyes graded by each observer compared with 0.42 between the 2 observers using a direct ophthalmoscope reported by Wong and White (17). It should be noted that direct ophthalmoscopic SVP grading by 2 assessors was not recorded for the patients in our study, so direct comparison between the inter-rater reliability of Spectralis video assessments and of direct ophthalmoscopy is not possible. The Spectralis and equivalent OCT systems are found in most neuro-ophthalmology clinics; thus, the technology for IR video SVP assessment is already widely available. Infrared video assessment of SVP in both eyes takes under a minute and can be used in both ambulatory and wheelchair-bound patients. Our group is undertaking further validation work using the infrared video assessment technique and SVP grading system described in this article to examine patients with suspected elevated ICP while undergoing continuous transcranial pressure measurements. The main limitation of the technique described is that the Spectralis OCT can only be used on subjects in a sitting position and, therefore, cannot be used to screen acutely ill bedbound patients such as those with traumatic brain injury. Further work is needed to develop this technique for use in patients who cannot be examined in a sitting position. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. STATEMENT OF AUTHORSHIP Category 1: a. Conception and design: J. A. McHugh, L. D'Antona, A. K. Toma, and F. D. Bremner; b. Acquisition of data: J. A. McHugh and F. D. Bremner; c. Analysis and interpretation of data: J. A. McHugh, L. D'Antona, A. K. Toma, and F. D. Bremner. Category 2: a. Drafting the manuscript: J. A. McHugh, L. D'Antona, and F. D. Bremner; b. Revising it for intellectual content: J. A. McHugh, L. D'Antona, A. K. Toma, and F. D. Bremner. Category 3: a. Final approval of the completed manuscript: J. A. McHugh, L. D'Antona, A. K. Toma, and F. D. Bremner. 15. 16. 17. 18. 19. REFERENCES 1. Berdahl JP, Yu DY, Morgan WH. The translaminar pressure gradient in sustained zero gravity, idiopathic intracranial McHugh et al: J Neuro-Ophthalmol 2020; 40: 174-177 20. hypertension, and glaucoma. Med Hypotheses. 2012;79:719- 724. Anderson RC, Kan P, Klimo P, Brockmeyer DL, Walker ML, Kestle JR. Complications of intracranial pressure monitoring in children with head trauma. J Neurosurg. 2004;101:53-58. Evans RW. Complications of lumbar puncture. Neurol Clin. 1998;16:83-105. Mizutani T, Manaka S, Tsutsumi H. Estimation of intracranial pressure using computed tomography scan findings in patients with severe head injury. Surg Neurol. 1990;33:178-184. Watanabe A, Kinouchi H, Horikoshi T, Uchida M, Ishigame K. Effect of intracranial pressure on the diameter of the optic nerve sheath. J Neurosurg. 2008;109:255-258. Hansen HC, Helmke K. The subarachnoid space surrounding the optic nerves. An ultrasound study of the optic nerve sheath. Surg Radiol Anat. 1996;18:323-328. Saqqur M, Zygun D, Demchuk A. Role of transcranial Doppler in neurocritical care. Crit Care Med. 2007;35:S216-S223. Shimbles S, Dodd C, Banister K, Mendelow AD, Chambers IR. Clinical comparison of tympanic membrane displacement with invasive ICP measurements. Acta Neurochir Suppl. 2005;95:197-199. Büki B, Avan P, Lemaire JJ, Dordain M, Chazal J, Ribári O. Otoacoustic emissions: a new tool for monitoring intracranial pressure changes through stapes displacements. Hear Res. 1996;94:125-139. Chen JW, Gombart ZJ, Rogers S, Gardiner SK, Cecil S, Bullock RM. Pupillary activity as an early indicator of elevated intracranial pressure: the Introduction of the Neurological Pupil Index. Surg Neurol Int. 2011;2:82. Kolecki R, Dammavalam V, Bin Zahid A, Hubbard M, Choudhry O, Reyes M, Han B, Wang T, Papas PV, Adem A, North E, Gilbertson DT, Kondziolka D, Huang JH, Huang PP, Samadani U. Elevated intracranial pressure and reversible eye-tracking changes detected while viewing a film clip. J Neurosurg. 2018;128:811-818. Steffen H, Eifert B, Aschoff A, Kolling GH, Völcker HE. The diagnostic value of optic disc evaluation in acute elevated intracranial pressure. Ophthalmology. 1996;103:1229-1232. Gibbs FA. Relationship between the pressure in the veins on the nerve head and the cerebrospinal fluid pressure. Arch Neurol Psychiatry. 1936;35:292-295. Morgan WH, Hazelton ML, Yu D-Y. Retinal venous pulsation: expanding our understanding and use of this enigmatic phenomenon. Prog Retin Eye Res. 2016;55:82-107. Levine DN, Bebie H. Phase and amplitude of spontaneous retinal vein pulsations: an extended constant inflow and variable outflow model. Microvasc Res. 2016;106:67-79. Levin BE. The clinical significance of spontaneous pulsations of the retinal vein. Arch Neurol. 1978;35:37-40. Wong SH, White RP. The clinical validity of the spontaneous retinal venous pulsation. J Neuroophthalmol. 2013;33:17-20. Seo JH, Kim TW, Weinreb RN, Kim YA, Kim M. Relationship of intraocular pressure and frequency of spontaneous retinal venous pulsation in primary open-angle glaucoma. Ophthalmology. 2012;119:2254-2260. Lee E, Kim TW, Kim JA, Kim JA, Kim H. Spontaneous retinal venous pulsation in unilateral primary open-angle glaucoma with low intraocular pressure. J Glaucoma. 2017;26:896-901. Hedges TR, Jr, Baron EM, Hedges TR III, Sinclair SH. The retinal venous pulsation. Ophthalmology. 1994;101:542-547. 177 Copyright © North American Neuro-Ophthalmology Society. Unauthorized reproduction of this article is prohibited. |
Date | 2020-06 |
Language | eng |
Format | application/pdf |
Type | Text |
Publication Type | Journal Article |
Source | Journal of Neuro-Ophthalmology, June 2020, Volume 40, Issue 2 |
Collection | Neuro-Ophthalmology Virtual Education Library - Journal of Neuro-Ophthalmology Archives: https://novel.utah.edu/jno/ |
Publisher | Lippincott, Williams & Wilkins |
Holding Institution | Spencer S. Eccles Health Sciences Library, University of Utah, 10 N 1900 E SLC, UT 84112-5890 |
Rights Management | © North American Neuro-Ophthalmology Society |
ARK | ark:/87278/s61z9ts4 |
Setname | ehsl_novel_jno |
ID | 1592880 |
Reference URL | https://collections.lib.utah.edu/ark:/87278/s61z9ts4 |