Description Over the past four decades, Multielectrode Array (MEA) devices have played a major role in electrophysiology by providing a simpler solution to simultaneous multi-site chronic extracellular recording: in vivo and in vitro. While a wide range of devices have been developed, almost all of them are limited to culturing and recording from one cell type, in vitro; and tissue surfaces, in vivo and in vitro. Most tissues are formed by different cell types that interact to maintain tissue function, like the heart which is composed mainly of cardio-myocytes and fibroblasts. Direct recording from such organs usually employs plunge-type electrodes which induce tissue damage and require better handling for sustenance. To better understand the functioning of such tissues, it is imperative to utilize recording systems that allow interactions between two or more cell types and at the same time sustain cultures with controlled cell number and distribution. In this thesis, the design, fabrication process, and characterization of an MEA device called the PerFlexMEA (Perforated Flexible MEA) is presented. It enables the generation and sustenance of a preparation with two cell types while recording their electrical activity. PerFlexMEA was developed using a thin (9?m) perforated Polycarbonate Track Etch (PCTE) membrane (3?m diam. pores, 200,000 pores/cm2) as substrate where cells can be cultured on both sides, allowing gap junction formation across the membrane via the pores. Cell number and distribution can be controlled on either side. The PerFlexMEA comprises a 4 × 5 array of square gold electrodes, each measuring 50 ?m × 50 ?m spaced 500 ?m apart. Parylene was patterned to insulate the leads (50 ?m thick) connecting the recording electrodes to the contact pads. A coinshaped device was designed to house the PerFlexMEA and to insulate its cell culture zone (wet) from contact pads (dry). Cardiomyocytes, isolated from neonatal mice were plated on the recording side of PerFlexMEA and electrical activity was recorded at a signal to noise ratio of 8.6 and peak to peak voltage of 200 ?V.
Type Text
Publisher University of Utah
Subject Excitable cells; Flexible; Multielectrode array; Perforated; Polycarbonate; PerFlexMEA
Dissertation Institution University of Utah
Dissertation Name Master of Science
Language eng
Rights Management Copyright © Abhijit Mondal 2011
Format application/pdf
Format Medium application/pdf
Format Extent 7,158,750 bytes
Identifier us-etd3,53672
Source Original housed in Marriott Library Special Collections, QP6.5 2011 .M66
ARK ark:/87278/s6xd1gf1
Setname ir_etd
ID 194585
Reference URL https://collections.lib.utah.edu/ark:/87278/s6xd1gf1
Back to Search Results